Key result
Individualized EV-ICD reprogramming safely prevents electromagnetic interference during welding after on-site testing.
Why the study?
The introduction of the Medtronic Aurora EV-ICD raises new questions regarding occupational safety, such as welding, for which no definitive guidelines or consensus exist.
Does tailored on-site electromagnetic interference testing and device reprogramming prevent inappropriate EMI detection from welding equipment in a patient with an extravascular ICD?
Case Report (n=1)
No
Does tailored on-site electromagnetic interference testing and device reprogramming prevent inappropriate EMI detection from welding equipment in a patient with an extravascular ICD?
Individualized on-site electromagnetic interference testing and device reprogramming can safely enable patients with an extravascular ICD to resume high-EMI occupational activities such as welding.
Answers are grounded in this paper and cite their sources.
Individualized on-site EMI testing may permit return to welding after EV-ICD implantation; leaves open generalizability beyond this single case.
No abstract yet. The journal hasn't published one we can show.
No takes yet. Share an insight, caveat, or question.
Römers et al. (2025) conducted a case report in Brugada syndrome type 2 with extravascular ICD (n=1). Welding equipment exposure was evaluated on Detection of electromagnetic interference (EMI) during welding. On-site testing and individualized device reprogramming safely allowed a patient with an extravascular ICD to resume welding without inappropriate detection of electromagnetic interference.