This case highlights the risk of inappropriate ICD shocks triggered by interrogation when the magnet switch is deactivated, emphasizing the need for multiple device inactivation methods.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to interrogation-triggered ICD shocks with deactivated magnet; case report leaves open need for multiple inactivation methods.
A 57-year-old man suffered multiple inappropriate shocks from an implantable cardioverter defibrillator that could not be inactivated because the magnet switch feature was deactivated. Attempts to interrogate the defibrillator caused shocks to be delivered. Emergent explantation was performed and a loose sensing lead was discovered. Sensing and defibrillation leads functioned normally, and testing of the explanted generator demonstrated no abnormalities. A new generator was put in place and the patient has been without shocks for 6 months. This case illustrates the need for robust methods of attenuating electromagnetic interference and the importance of multiple methods of device inactivation.