Why the study?
Does exposure to electronic surveillance systems cause inappropriate shocks in patients with ICDs?
Does exposure to electronic surveillance systems cause inappropriate shocks in patients with ICDs?
Electronic surveillance systems can cause inappropriate shocks in patients with ICDs due to electromagnetic interference.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for EMI-induced inappropriate ICD shocks from surveillance systems; larger studies needed to quantify risk.
We present a case report of an inappropriate discharge from a fourth generation implantable cardioverter defibrillator (ICD) as a result of exposure to electromagnetic interference. A 60-year-old man implanted with a Medtronic 7219D defibrillator experienced shocks without preceding symptoms while walking through an electronic surveillance system in a store. Though this has been reported, the mechanism of the interaction remains unexplained. The electrogram storage capabilities of this particular device enabled us to establish that this resulted from inappropriate sensing of the electromagnetic interference.