Key result
A defibrillator shock lead that had been in place in the coronary sinus for almost 10 years was successfully removed using laser lead extraction.
Case Report (n=1)
Demonstrates the feasibility of percutaneous laser lead extraction for a coronary sinus defibrillator lead that has been in place for nearly a decade.
Answers are grounded in this paper and cite their sources.
Supports feasibility of laser extraction for chronic coronary sinus defibrillator leads; hypothesis-generating and should not yet change practice.
Placement of leads into the coronary sinus (CS) is now routinely used for cardiac resynchronization therapy. However, the CS can also be used for the placement of defibrillator shock leads. There is evidence demonstrating that CS leads in situ for the short and medium term can be successfully removed utilizing percutaneous methods. We report a case of CS laser lead extraction where a defibrillator lead had been in place for almost 10 years.
No takes yet. Share an insight, caveat, or question.
Hamid et al. (2009) conducted a case report in Defibrillator lead in coronary sinus (n=1). Coronary sinus laser lead extraction was evaluated on Successful lead extraction. A defibrillator shock lead that had been in place in the coronary sinus for almost 10 years was successfully removed using laser lead extraction.