Key result
Coronary artery spasm caused ST segment elevation that led to QRS double counting and an inappropriate ICD shock in a patient with unexplained syncope.
Case Report (n=1)
Coronary artery spasm can cause ST segment elevation that may be oversensed by an ICD (double counting of QRS), leading to inappropriate shocks.
Answers are grounded in this paper and cite their sources.
Raises caution for spasm-induced oversensing in ICD patients with syncope; leaves open prevalence and management implications.
Coronary artery spasm can cause both brady- and tachyarrhythmia, through induction of AV block (usually linked to coronary spasm of the right coronary artery) or ventricular tachycardia/fibrillation linked to extensive myocardial ischemia. The electrocardiographic aspect of coronary artery spasm is an ST segment elevation. We describe the case of patient implanted with an implantable cardioverter defibrillator (ICD) for unexplained syncope which, during coronary artery spasm, received an inappropriate device firing due to ST segment elevation, leading to a double count of the QRS by the ICD.
No takes yet. Share an insight, caveat, or question.
Mascioli et al. (2007) conducted a case report in Coronary artery spasm in a patient with an ICD (n=1). Coronary artery spasm was evaluated on Inappropriate ICD firing. Coronary artery spasm caused ST segment elevation that led to QRS double counting and an inappropriate ICD shock in a patient with unexplained syncope.