Key result
Transjugular lead fragment extraction and totally subcutaneous ICD implantation were performed in a 28-year-old man with severe tricuspid regurgitation and ICD lead failure.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
This case report describes the use of transjugular lead fragment extraction to address severe tricuspid regurgitation in a patient with a failed ICD lead.
A 28-year-old-man with a history of human immunodeficiency virus (HIV), end-stage renal disease requiring hemodialysis, and HIV-related cardiomyopathy with a severely reduced left ventricular ejection fraction (LVEF) and severe tricuspid regurgitation (TR) per a transthoracic echocardiogram (TTE) underwent implantation of a single-chamber implantable cardioverter-defibrillator (ICD) in 2009 (Riata 1581, St Jude Medical, St Paul, MN). Because of lead failure resulting in electrical noise manifesting as nonphysiological signals, a totally subcutaneous ICD (Boston Scientific, Inc, Natick, MA) was implanted.
No takes yet. Share an insight, caveat, or question.
Schaller et al. (2015) conducted a case report in HIV-related cardiomyopathy, severe tricuspid regurgitation, and ICD lead failure (n=1). Transjugular lead fragment extraction and totally subcutaneous ICD implantation was evaluated. Transjugular lead fragment extraction and totally subcutaneous ICD implantation were performed in a 28-year-old man with severe tricuspid regurgitation and ICD lead failure.
Synapse has enriched 10 related papers. Consider them for comparative context: