Key result
A dual-modality volume measurement combining ultrasound and electric impedance achieved limits of agreement of -0.83 ± 1.54%, outperforming ultrasound and impedance.
Why the study?
Real-time volume measurement in VADs could enable physiological control, but single-sensor modalities require re-calibration during the critical post-implantation period.
A dual-modality ultrasound and electric impedance sensor integrated into a VAD cannula provides accurate ventricular volume measurements in vitro without the need for recalibration.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for dual-modality VAD volume sensing; prospective human studies needed before clinical use.
OBJECTIVE: Ventricular assist devices (VADs) are implanted in patients suffering from end-stage heart failure to sustain the blood circulation. Real-time volume measurement could be a valuable tool to monitor patients and enable physiological control strategies to provide individualized therapy. However, volume measurement using one sensor modality requires re-calibration in the critical time post VAD implantation. METHODS: To overcome this limitation, we have integrated ultrasound and impedance volume measurement techniques into a cannula of an apical VAD. We tested both modalities across a volume range from 140-420 mL using two differently sized and shaped biventricular silicon heart phantoms, which were subjected to physiological pressures in an in-vitro test bench. We compared results from standard calibrated measurements with calculations found by a quadratic optimization for the single modality and their combination (dual-modality) and validated the results using twofold cross-validation. RESULTS: The dual-modality approach resulted in most favorable limits of agreement (LOA) of -0.83 ± 1.54% compared to -13.88 ± 5.90% for ultrasound and -43.45 ± 10.28% for electric impedance, separately. CONCLUSION: The results of the dual-modality approach were as accurate as the standard calibrated measurement and valid over a large range of volumes (140-420 mL). In this in-vitro study, we show how a dual-modality ventricular volume measurement of ultrasound and electric impedance increases the robustness and renders calibration obsolete. SIGNIFICANCE: Ventricular volumes could be measured accurately in the critical period post VAD implantation despite ventricular remodeling.
No takes yet. Share an insight, caveat, or question.
A 2021 study studied End-stage heart failure (in-vitro model). Dual-modality volume measurement (ultrasound and electric impedance) vs. Single modality (ultrasound or electric impedance) was evaluated on Limits of agreement (LOA) for volume measurement. A dual-modality volume measurement combining ultrasound and electric impedance achieved limits of agreement of -0.83 ± 1.54%, outperforming ultrasound and impedance.