Key result
The cumulative number of noncontact catheter electrodes delivering pulsed field energy correlated with postoperative hemolysis markers, including lactate dehydrogenase.
Why the study?
It remains unclear whether pulsed field ablation applications delivered from tissue proximity indication-negative (noncontact) variable-loop circular catheter electrodes contribute to hemolysis.
Is pulsed field ablation delivered from noncontact catheter electrodes associated with postprocedural hemolysis in patients undergoing their first PFA?
Cohort (n=20)
No
Is pulsed field ablation delivered from noncontact catheter electrodes associated with postprocedural hemolysis in patients undergoing their first PFA?
Delivering pulsed field ablation energy from noncontact catheter electrodes is associated with increased postprocedural hemolysis, suggesting that minimizing noncontact applications may improve procedural safety.
Answers are grounded in this paper and cite their sources.
Minimizing noncontact VLCC applications may reduce hemolysis risk; supports hypothesis-generating association in this small observational cohort.
Background: Pulsed field ablation (PFA) for atrial fibrillation can induce hemolysis, particularly when pulsed field (PF) energy is delivered from catheter electrodes that are not in contact with myocardial tissue. The variable-loop circular catheter (VLCC) incorporates tissue proximity indication (TPI) software to identify catheter-tissue contact. Whether PFA applications delivered from TPI-negative (noncontact) VLCC electrodes contribute to hemolysis remains unclear. Objective: The purpose of this study was to determine whether PFA delivered from TPI-negative VLCC electrodes is associated with postprocedural hemolysis. Methods: We retrospectively analyzed 20 consecutive patients who underwent their first PFA using the VLCC and CARTO 3 system (Biosense Webster Inc.). For each 3-pulse application set, catheter-tissue contact was assessed using TPI. We counted the cumulative number of VLCC electrode uses delivering PF energy; among these, those that were TPI-negative were summed and defined as the cumulative number of TPI-negative electrode uses (nTPI-EU). Postoperative lactate dehydrogenase (LDH), total bilirubin (T-Bil), and the postoperative-to-preoperative haptoglobin ratio served as hemolysis markers. Conclusion: The cumulative number of noncontact catheter electrodes delivering PF energy was associated with hemolysis. Strategies that minimize the number of noncontact catheter electrodes, in addition to limiting total PFA applications, may reduce the risk of hemolysis.
No takes yet. Share an insight, caveat, or question.
Yoshimura et al. (2025) conducted a cohort in Atrial fibrillation (n=20). Pulsed field ablation from noncontact (TPI-negative) catheter electrodes was evaluated on Correlation between cumulative number of TPI-negative electrode uses (nTPI-EU) and postoperative hemolysis markers (r = 0.663 for LDH; r = 0.736 for T-Bil, p=<0.01). The cumulative number of noncontact catheter electrodes delivering pulsed field energy correlated with postoperative hemolysis markers, including lactate dehydrogenase.