Key result
PFA with TPI-ON mode reduces hemolysis vs TPI-OFF in pigs while preserving acute transmural lesions.
Why the study?
No clinical studies have reported on the correlation between activation of the tissue proximity-indicating (TPI) function and hemolysis during pulsed field ablation for atrial fibrillation.
Does pulsed field ablation in TPI-ON mode reduce hemolysis compared to TPI-OFF mode in a porcine atrial myocardium model?
Does pulsed field ablation in TPI-ON mode reduce hemolysis compared to TPI-OFF mode in a porcine atrial myocardium model?
The use of tissue proximity-indicating (TPI) function during pulsed field ablation in a preclinical model reduces in vitro hemolysis while maintaining effective transmural lesion formation.
Answers are grounded in this paper and cite their sources.
Supports clinical testing of TPI-ON pulsed field ablation to limit hemolysis; hypothesis-generating in porcine model.
BACKGROUND: In pulsed field ablation (PFA), adequate catheter-tissue contact is crucial as it can effectively reduce the occurrence of hemolysis. Activation of the tissue proximity-indicating (TPI) function ensures effective catheter-tissue apposition. However, to date, no relevant clinical studies have been reported on the correlation between TPI function activation and hemolysis during PFA for atrial fibrillation (AF). OBJECTIVES: To investigate in depth the correlation between the "TPI" function of catheters and postoperative hemolysis following pulsed field ablation (PFA) for atrial fibrillation (AF) from hematological and pathological perspectives. METHODS: By applying PFA treatment in both TPI-ON and TPI-OFF modes to isolated porcine atrial myocardial tissue and in situ atrial myocardium from six pigs, the effects on hemolysis were observed under these two scenarios. RESULTS: In vitro, plasma free hemoglobin (fHb) increased with the number of applications and was lower with TPI-ON than with TPI-OFF at matched application numbers (p < 0.05). In vivo, both groups achieved transmural lesions with comparable lesion depth. Mean fHb values at 5 and 30 min were numerically lower with TPI-ON, but the group-by-time interaction was not significant (p = 0.153). CK and CK-MB increased transiently within 24 h, whereas urea and creatinine showed no consistent between-group pattern; none of the four biochemical group-by-time interactions remained significant after Holm correction. Renal histology showed no irreversible parenchymal injury in either group. CONCLUSION: In this preclinical model, use of the impedance-based TPI function was associated with lower post-PFA hemolysis while preserving acute transmural lesion formation. These findings support further validation of TPI-guided catheter positioning in clinical studies.
No takes yet. Share an insight, caveat, or question.
Yan et al. (2026) studied Atrial fibrillation (n=6). Pulsed field ablation in TPI-ON mode vs. Pulsed field ablation in TPI-OFF mode was evaluated on Plasma free hemoglobin (fHb) levels (p=< 0.05). In a preclinical porcine model, pulsed field ablation using TPI-ON mode was associated with lower in vitro hemolysis compared to TPI-OFF while preserving acute transmural lesions.
Synapse has enriched 10 related papers. Consider them for comparative context: