Key result
Olive-configured pulsed field ablation linked to greater hemolysis than basket or flower configurations.
Why the study?
Pulsed field ablation (PFA) for atrial fibrillation is associated with hemolysis, and it was unclear how variations in pentaspline catheter configuration influence hemolysis intensity.
Does the olive catheter configuration increase hemolysis compared to the flower or basket configuration during pulsed field ablation for atrial fibrillation?
Observational (n=20)
Does the olive catheter configuration increase hemolysis compared to the flower or basket configuration during pulsed field ablation for atrial fibrillation?
Pulsed field ablation using a pentaspline catheter in the olive configuration induces significantly greater hemolysis both in vitro and in vivo compared to flower or basket configurations.
Answers are grounded in this paper and cite their sources.
Olive configuration was associated with greater hemolysis during PFA; leaves open optimal positioning to minimize erythrocyte damage in future trials.
Background Pulsed field ablation (PFA) is an emerging ablation technology for atrial fibrillation (AF) and associated with hemolysis. Objective This study seeks to evaluate how variations in catheter configuration could influence hemolysis intensity. Methods: Using a translational approach, hemolysis induced by a pentaspline PFA catheter was assessed in an in vitro erythrocyte concentrate model. PFA applications in flower, basket or olive configuration were delivered in erythrocyte concentrate, and hemolysis index was used to measure hemolysis. In twenty patients who underwent pulmonary vein isolation with PFA for AF using the pentaspline catheter hemolysis markers were measured before and immediately after the procedure and on the first and second post-procedural days. Results In vitro application in olive configuration led to significantly more hemolysis in erythrocyte concentrate than basket or flower configuration. After 20 applications, hemolysis index increased by 44 in the olive vs. 24 in the flower and 22.5 in the basket group. In vivo , hemolysis index increased significantly from 6.7 ± 11.77 before to 24.65 ± 6.28 immediately after PFA treatment in patients (p<0.0001). Haptoglobin decreased significantly from 1 ± 0.37 g/l before procedure to 0.53 ± 0.36 g/l on the first post-procedural day. Patients treated with olive applications demonstrated a significantly greater decrease in haptoglobin (-0.38 g/l vs. -0.55 g/l; p=0.02). No case of acute kidney injury occurred. Conclusion This study demonstrates that PFA applications with different pentaspline catheter configurations are associated with varying extents of hemolysis. The olive configuration was associated with more hemolysis than basket or flower configuration.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2026) conducted an observational in Atrial fibrillation (n=20). Pulsed field ablation with pentaspline catheter in olive configuration vs. Basket or flower catheter configuration was evaluated on Decrease in haptoglobin on the first post-procedural day (p=0.02). Pulsed field ablation in the olive catheter configuration caused a greater haptoglobin decrease in vivo and more in vitro hemolysis than basket or flower configurations.
Synapse has enriched 10 related papers. Consider them for comparative context: