Key result
Prepectoral and submuscular ICD implantation techniques yielded similar defibrillation thresholds (averaging 9-10 J in both groups) and pacing thresholds, with 0 complications or operative deaths.
Why the study?
Does prepectoral ICD implantation compared to submuscular implantation affect defibrillation thresholds or complications in patients receiving transvenous ICDs?
Observational (n=45)
Does prepectoral ICD implantation compared to submuscular implantation affect defibrillation thresholds or complications in patients receiving transvenous ICDs?
Both prepectoral and submuscular ICD implantation techniques are safe and effective, with similar defibrillation and pacing thresholds, allowing tailoring of the approach based on patient subcutaneous tissue thickness.
Either technique appears safe with comparable thresholds; extends observational data but remains hypothesis-generating pending RCTs.
The purpose of this study was to compare the two techniques of pectoral ICD implantation, prepectoral and submuscular, performed by an electrophysiologist in the catheterization laboratory with use of general or local anesthesia in 45 consecutive patients. Over a period of 30 months, we implanted pectoral transvenous ICDs in 43 men and 2 women, aged 59 +/- 12 years, with use of general (n = 20) or local (n = 25) anesthesia in the catheterization laboratory. Patients had coronary (n = 30) or valvular (n = 4) disease, cardiomyopathy (n = 10) or no organic disease (n = 1), a mean left ventricular ejection fraction of 31%, and presented with ventricular tachycardia (n = 40) or fibrillation (n = 5). One-lead ICD systems (18 Endotak, 10 Transvene/8 Sprint, 2 EnGuard) were used in 38 patients, 2-lead (5 Transvene, 1 EnGuard) systems in 6 patients, and 1 atrioventricular lead ICD system in 1 patient. The prepectoral technique was employed in 29 patients with adequate subcutaneous tissue, while the submuscular technique was used in 16 patients who had a thin layer of subcutaneous tissue. The defibrillation threshold averaged 9-10 J in both groups and there were no differences in pace/sense thresholds. All implants were entirely transvenous with no subcutaneous patch. Biphasic ICD devices were employed in all patients. Active or hot can devices were used in 39 patients. There were no complications, operative deaths, or infections. Patients were discharged at a mean of 3 days. All devices functioned well at predis-charge testing. Over 14 +/- 8 months, 20 patients received appropriate device therapy (antitachycardia pacing or shocks). No late complications occurred. One patient died at 3 months of pump failure; there were no sudden deaths. In conclusion, for exclusive pectoral implantation of transvenous ICDs, electrophysiologists should master both prepectoral and submuscular techniques. One can thus avoid potential skin erosion or need for abdominal implantation in patients with a thin layer of subcutaneous tissue. Finally, there are no differences in pacing or defibrillation thresholds between the two techniques.
No takes yet. Share an insight, caveat, or question.
Manolis et al. (1999) conducted an observational in Ventricular tachycardia or fibrillation requiring ICD (n=45). Prepectoral ICD implantation technique vs. Submuscular ICD implantation technique was evaluated on Defibrillation and pacing/sensing thresholds, and complications. Prepectoral and submuscular ICD implantation techniques yielded similar defibrillation thresholds (averaging 9-10 J in both groups) and pacing thresholds, with 0 complications or operative deaths.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: