Key result
The Epic mitral valve demonstrated stable hemodynamics over 5 years with a low 5% incidence of reintervention due to prosthesis failure at a median follow-up of 3.0 years.
Why the study?
Long-term echocardiographic data on the Epic mitral valve porcine xenograft bioprosthesis are lacking, and postreintervention outcomes of failed Epic valves are unknown.
What are the long-term echocardiographic outcomes, mechanisms of failure, and reintervention outcomes of the Epic porcine xenograft bioprosthesis in the mitral position?
Cohort (n=1,397)
No
What are the long-term echocardiographic outcomes, mechanisms of failure, and reintervention outcomes of the Epic porcine xenograft bioprosthesis in the mitral position?
The Epic mitral valve bioprosthesis demonstrates stable hemodynamics over 5 years with a low incidence of structural valve deterioration and reintervention at 10 years.
Supports mid-term Epic mitral durability in observational data; leaves open need for randomized long-term comparisons.
OBJECTIVES: Long-term echocardiographic reports on mitral valve (MV) porcine xenograft bioprosthesis (Epic) are lacking, and postreintervention outcomes of failed Epic are unknown. We aimed to assess the mechanisms and independent predictors of Epic failures and to compare short- and mid-term outcomes according to reintervention type. METHODS: We included consecutive patients (n = 1397; mean age: 72 ± 8 years; 46% female; mean follow-up: 4.8 years) who received the Epic during mitral valve replacement (MVR) at our institution. Clinical, echocardiographic, reintervention, and outcomes data were retrieved from our prospective institution's database or government statistics. RESULTS: Gradients and effective orifice area of the Epic were stable over 5-years follow-up. A total of 70 (5%) patients had a MV reintervention at median follow-up of 3.0 (0.7-5.4) years due to prosthesis failure, by redo-MVR (n = 38; 54%), valve-in-valve (n = 19; 27%), paravalvular leak (PVL) closure (n = 12; 17%), or thrombectomy (n = 1). Mechanisms of failure were 27 (1.9%) structural valve deterioration (SVD; all leaflet tear); 16 (1.1%) non-SVD (15 PVL, 1 pannus); 24 (1.7%) endocarditis; and 4 (0.3%) thrombosis. Freedom from all-cause and SVD-related MV reintervention at 10 years are 88% and 92%, respectively. Independent predictors of reintervention were age, baseline atrial fibrillation, initial MV etiology, and moderate or greater PVL at discharge (all P ≤ .05). Comparison of redo-MVR and valve-in-valve revealed no significant difference in early outcomes or mid-term mortality (all P ≥ .16). CONCLUSIONS: The Epic Mitral valve has stable hemodynamics through 5 years and is associated with low incidence of SVD and reintervention, mostly due to endocarditis and leaflet tear without calcification. Reintervention type had no influence on early outcomes and mid-term mortality.
No takes yet. Share an insight, caveat, or question.
Bernard et al. (2023) conducted a cohort in Mitral valve replacement (n=1,397). Epic porcine xenograft bioprosthesis was evaluated on Mitral valve reintervention due to prosthesis failure. The Epic mitral valve demonstrated stable hemodynamics over 5 years with a low 5% incidence of reintervention due to prosthesis failure at a median follow-up of 3.0 years.