Key result
Higher operator experience was associated with better leadless pacemaker implantation quality and increased procedural efficiency.
Why the study?
Recent real-world data raised concerns regarding the learning curve of new operators and implantation quality during leadless pacemaker implantation.
Does operator experience improve implantation quality and procedural efficiency in patients undergoing leadless pacemaker implantation?
Observational (n=111)
Yes
Does operator experience improve implantation quality and procedural efficiency in patients undergoing leadless pacemaker implantation?
Operator experience is significantly associated with higher implantation quality and procedural efficiency during leadless pacemaker implantation, highlighting the importance of the learning curve.
Answers are grounded in this paper and cite their sources.
Supports learning-curve effects in leadless pacing; hypothesis-generating for experience thresholds before practice change.
AIMS: Leadless cardiac pacemaker (PM) implantation differs from conventional PM implantation. While the procedure has been considered safe, recent real-world data raised concerns about the learning curve of new operators and their implantation quality. The goal of this study was to investigate the influence of the first operator's experience on leadless PM implantation quality and procedural efficiency. METHODS AND RESULTS: We performed a bicentric analysis of all Micra TPS™ implantations in two large tertiary referral hospitals. We assessed both leadless PM implantation quality based on the absence of complications (requiring intervention or prolonged hospitalization), good electrical performance (pacing threshold ≤ 1.5 V/0.24 ms, R-wave amplitude > 5 mV), and acceptable fluoroscopy duration (<10 min) as well as procedural efficiency in relation to the operator's experience. Univariate and multivariate logistic regression analyses were performed to identify predictors for implantation quality and procedural efficiency. Leadless PM implantation was successful in 106/111 cases (95.5%). Three patients (2.7%) experienced acute complications (one cardiac tamponade, one femoral bleeding, one posture-related PM exit block). Multivariate analysis showed that implantation quality of more experienced first operators was higher [odds ratio 1.09 (95% confidence interval 1.00-1.19), P = 0.05]. Procedural efficiency increased with operator experience as evidenced by an inverse correlation of procedure time, time to the first deployment, fluoroscopy time, and the number of procedures performed (all P < 0.05). CONCLUSION: The operator's learning curve is a critical factor for leadless PM implantation quality and procedural efficiency.
No takes yet. Share an insight, caveat, or question.
Haeberlin et al. (2020) conducted an observational in Leadless cardiac pacemaker implantation (n=111). Operator experience vs. Less operator experience was evaluated on Implantation quality (absence of complications, good electrical performance, acceptable fluoroscopy duration) (OR 1.09, 95% CI 1.00-1.19, p=0.05). Higher operator experience was associated with better leadless pacemaker implantation quality and increased procedural efficiency.
Synapse has enriched 10 related papers. Consider them for comparative context: