Key result
Intravascular measurements of pressure and flow in the rabbit aorta proved feasible, though simultaneous measurement reduced absolute value reliability by reducing the cross-sectional diameter.
Intravascular measurement of pressure and flow in rabbits is feasible for parameter fitting in digital hemodynamic models, provided catheter size is carefully selected to avoid obstructing flow.
Answers are grounded in this paper and cite their sources.
Supports rabbit aortic measurements for digital hemodynamic model fitting; leaves open translation to larger models or humans.
The objective was to describe and evaluate methods of in vivo intravascular assessment of pressure and flow within the systemic arterial tree for further digital processing and individualized simulation. In rabbits intra-aortic pressure and flow were measured by insertion of microcatheters and/or a Doppler probe, which were moved to different angiographically documented sites. Various catheters and sheaths were evaluated prior to digitized processing of the obtained data, which was based on electrotechnical analoga for description of elastic vascular segments. Fourier transformation and filtering provided signal data for further identification of the investigated parameters. Catheters with an inner diameter > or = 0.9 mm produced reliable pressure signals. Simultaneous measurement of pressure and flow reduced the reliability of the absolute values by reduction of the cross-sectional diameter of the investigated vascular segments. Numerical optimization methods were used to identify parameters in the proposed models from the measured data sets. Characterization of hemodynamics by intravascular measurements of pressure and flow proved feasible. Pressure recordings should be judged critically if the cross-sectional area of the catheter covers a significant fraction of the vessel. Post-processing of such data sets may assist in future treatment planning and simulation of concomitant changes of intravascular pressure and flow.
No takes yet. Share an insight, caveat, or question.
Frings-Naberschulte et al. (2002) studied this question. Intravascular measurement of pressure and flow was evaluated on Feasibility and reliability of intravascular measurements of pressure and flow. Intravascular measurements of pressure and flow in the rabbit aorta proved feasible, though simultaneous measurement reduced absolute value reliability by reducing the cross-sectional diameter.
Synapse has enriched 10 related papers. Consider them for comparative context: