Does sequential ventricular defibrillation using monophasic double-pulses of trapezoidal shape achieve successful defibrillation in open-chest dogs with experimental ventricular fibrillation?
Does sequential ventricular defibrillation using monophasic double-pulses of trapezoidal shape achieve successful defibrillation in open-chest dogs with experimental ventricular fibrillation?
Sequential ventricular defibrillation using monophasic double-pulses achieved an 80% success rate at low energy (2.4 J) in an experimental open-chest dog model.
Answers are grounded in this paper and cite their sources.
May inform low-energy defibrillator design; leaves open translation from this canine model to clinical use.
A method of sequential ventricular defibrillation using monophasic double-pulses of trapezoidal shape and of variable voltage, width, and separation is described, and its portable apparatus presented. Results in experimental ventricular fibrillation in open-chest dogs showed an 80% success rate using the two pulses at a critical separation of 100 msec and an average total energy of 2·4 J.