Key result
In postoperative cardiac surgical patients, portable bedside tests for prothrombin time were clinically accurate and precise compared to standard laboratory tests, whereas aPTT tests were not.
Why the study?
Are portable bedside coagulation monitoring tests accurate and precise compared to standard laboratory tests in patients after cardiac surgery?
Observational (n=141)
Are portable bedside coagulation monitoring tests accurate and precise compared to standard laboratory tests in patients after cardiac surgery?
Bedside PT testing is accurate and precise in postoperative cardiac surgical patients, but bedside aPTT testing is not clinically acceptable compared to standard laboratory tests.
May support bedside PT monitoring post-cardiac surgery; leaves open prospective validation before any practice change.
We compared portable bedside tests of whole blood coagulation with standard laboratory plasma coagulation tests to assess the accuracy and precision of the bedside tests in a clinical setting (postcardiac surgery). The Ciba Corning 512 Coagulation Monitor (Ciba Corning Diagnostics Corp., Medfield, MA) and the Hemochron 801 (International Technidyne Corp., Edison, NJ) were tested. One hundred forty-one patients who underwent cardiac surgery requiring cardiopulmonary bypass were evaluated upon arrival in the intensive care unit. Nine milliliters of fresh whole blood were used to obtain the prothrombin time (PT), activated partial thromboplastin time (aPTT), thrombin time (TT), and the heparin-neutralized thrombin time (HNTT) that were measured by bedside coagulation monitors. TT and HNTT were only measured by the Hemochron method. Blood from the same sample was also sent to the Hospital Stat Laboratory for simultaneous comparison of bedside results and standard coagulation tests. For PT and aPTT testing, an accuracy of +/- 10% and a precision of +/- 25% were considered clinically acceptable. Both of the PT tests met the dual criteria for clinical acceptability, but the aPTT tests did not meet either criterion. The difference between Hemochron TT and Hemochron HNTT correlated weakly, but significantly, with laboratory aPTT ratio (r = 0.52, P < 0.001). The slope of the regression line indicated that a TT-HNTT difference > 30 s correlated with an aPTT > 1.5 x control. We conclude that, in the postoperative cardiac surgical patient, PT was both accurate and precise in two commercially available tests, but aPTT was not.(ABSTRACT TRUNCATED AT 250 WORDS)
No takes yet. Share an insight, caveat, or question.
Reich et al. (1993) conducted an observational in Postcardiac surgery (n=141). Portable bedside tests of whole blood coagulation vs. Standard laboratory plasma coagulation tests was evaluated on Accuracy (+/- 10%) and precision (+/- 25%) of PT and aPTT testing. In postoperative cardiac surgical patients, portable bedside tests for prothrombin time were clinically accurate and precise compared to standard laboratory tests, whereas aPTT tests were not.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: