Key result
Vital signal detection using two adjacent radars at different frequencies in the 5.8 GHz band achieved a mean accuracy of 97.8% for a subject moving at a maximum velocity of 53.4 mm/s.
Why the study?
Vital signal detection using continuous-wave radar suffers from signal degradation caused by a subject's body movement.
A novel multi-radar system operating in the 5.8 GHz band can accurately detect vital signals such as respiration and heartbeat by mitigating the effects of body movement.
Multi-radar CW detection may reduce motion artifacts in vital signs; leaves open clinical validation before adoption.
Vital signal detection using multiple radars is proposed to reduce the signal degradation from a subject's body movement. The phase variation in the transceiving signals of continuous-wave radar due to respiration and heartbeat is generated by the body surface movement of the organs monitored in the line-of-sight (LOS) of the radar. The body movement signals obtained by two adjacent radars can be assumed to be the same over a certain distance. However, the vital signals are different in each radar, and each radar has a different LOS because of the asymmetric movement of lungs and heart. The proposed method uses two adjacent radars with different LOS to obtain correlated signals that reinforce the difference in the asymmetrical movement of the organs. The correlated signals can improve the signal-to-noise ratio in vital signal detection because of a reduction in the body movement effect. Two radars at different frequencies in the 5.8 GHz band are implemented to reduce direct signal coupling. Measurement results using the radars arranged at angles of 30°, 45°, and 60° showed that the proposed method can detect the vital signals with a mean accuracy of 97.8% for the subject moving at a maximum velocity of 53.4 mm/s.
No takes yet. Share an insight, caveat, or question.
Jang et al. (2021) studied Vital signal detection. Two adjacent radars at different frequencies in the 5.8 GHz band was evaluated on Mean accuracy of vital signal detection. Vital signal detection using two adjacent radars at different frequencies in the 5.8 GHz band achieved a mean accuracy of 97.8% for a subject moving at a maximum velocity of 53.4 mm/s.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: