Key result
In an in vitro model, reduced aortic root distensibility significantly increased the geometric orifice area and valve opening and closing velocities of a bioprosthetic aortic valve.
Why the study?
Does aortic root distensibility affect the haemodynamic parameters and valve kinematics of a bioprosthetic aortic valve in an in vitro model?
Does aortic root distensibility affect the haemodynamic parameters and valve kinematics of a bioprosthetic aortic valve in an in vitro model?
In an in vitro model, aortic root stiffness significantly altered the kinematics and geometric orifice area of a bioprosthetic aortic valve, which may impact mechanical load and long-term structural valve deterioration.
Reduced root distensibility alters bioprosthetic valve kinematics in vitro; hypothesis-generating for durability impact, needs in vivo confirmation.
Objectives: In this study, the influence of aortic root distensibility on the haemodynamic parameters and valve kinematics of a bioprosthetic aortic valve was investigated in a controlled in vitro experiment. Methods: An Edwards INTUITY Elite 21 mm sutureless aortic valve (Edwards Lifesciences, Irvine, CA, USA) was inserted in three transparent aortic root phantoms with different wall thicknesses (0.55, 0.85 and 1.50 mm) mimicking different physiological distensibilities. Haemodynamic measurements were performed in an in vitro flow loop at heart rates of 60, 80 and 100 bpm with corresponding cardiac outputs of 3.5, 4.0 and 5.0 l/min and aortic pressures of 100/60, 120/90 and 145/110 mmHg, respectively. Aortic valve kinematics were assessed using a high-speed camera. The geometric orifice area (GOA) was measured by counting pixels in the lumen of the open aortic valve. The effective orifice area (EOA) was calculated from the root-mean-square value of the systolic aortic valve flow rate and the mean systolic trans-valvular pressure gradient. Results: The tested aortic root phantoms reproduce physiological distensibilities of healthy individuals in age groups ranging from 40 to 70 years (±10 years). The haemodynamic results show only minor differences between the aortic root phantoms: the trans-valvular pressure gradient tends to increase for stiffer aortic roots, whereas the systolic aortic valve flow rate remains constant. As a consequence, the EOA decreased slightly for less distensible aortic roots. The GOA and the aortic valve opening and closing velocities increase significantly with reduced distensibility for all haemodynamic measurements. The resulting mean systolic flow velocity in the aortic valve orifice is lower for the stiffer aortic root. Conclusions: Aortic root distensibility may influence GOA and aortic valve kinematics, which affects the mechanical load on the aortic valve cusps. Whether these changes have a significant effect on the onset of structural valve deterioration of bioprosthetic heart valves needs to be further investigated.
No takes yet. Share an insight, caveat, or question.
Jahren et al. (2016) studied this question. Aortic root distensibility was evaluated on Haemodynamic parameters and valve kinematics. In an in vitro model, reduced aortic root distensibility significantly increased the geometric orifice area and valve opening and closing velocities of a bioprosthetic aortic valve.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: