Key result
Intrinsic cardiac gating achieved excellent agreement with ECG-derived gating, with a coverage probability for a difference <5 ms of 89.2% in mice and 85.9% in rats.
Intrinsic cardiac gating using raw CT projection data is a reliable alternative to external ECG gating for small-animal imaging, enabling high-quality functional assessment.
May enable ECG-free gating in rodent models; leaves open human validation before clinical translation.
BACKGROUND: A projection-based method of intrinsic cardiac gating in small-animal computed tomography imaging is presented. METHODS AND RESULTS: In this method, which operates without external ECG monitoring, the gating reference signal is derived from the raw data of the computed tomography projections. After filtering, the derived gating reference signal is used to rearrange the projection images retrospectively into data sets representing different time points in the cardiac cycle during expiration. These time-stamped projection images are then used for tomographic reconstruction of different phases of the cardiac cycle. Intrinsic gating was evaluated in mice and rats and compared with extrinsic retrospective gating. An excellent agreement was achieved between ECG-derived gating signal and self-gating signal (coverage probability for a difference between the 2 measurements to be less than 5 ms was 89.2% in mice and 85.9% in rats). Functional parameters (ventricular volumes and ejection fraction) obtained from the intrinsic and the extrinsic data sets were not significantly different. The ease of use and reliability of intrinsic gating were demonstrated via a chemical stress test on 2 mice, in which the system performed flawlessly despite an increased heart rate. Because of intrinsic gating, the image quality was improved to the extent that even the coronary arteries of mice could be visualized in vivo despite a heart rate approaching 430 bpm. Feasibility of intrinsic gating for functional imaging and assessment of cardiac wall motion abnormalities was successfully tested in a mouse model of myocardial infarction. CONCLUSIONS: Our results demonstrate that self-gating using advanced software postprocessing of projection data promises to be a valuable tool for rodent computed tomography imaging and renders ECG gating with external electrodes superfluous.
No takes yet. Share an insight, caveat, or question.
Dinkel et al. (2008) studied Small-animal computed tomography imaging. Intrinsic cardiac gating (projection-based method) vs. Extrinsic retrospective gating (ECG-derived) was evaluated on Coverage probability for a difference between the 2 measurements to be less than 5 ms. Intrinsic cardiac gating achieved excellent agreement with ECG-derived gating, with a coverage probability for a difference <5 ms of 89.2% in mice and 85.9% in rats.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: