Answers are grounded in this paper and cite their sources.
RATIONALE AND OBJECTIVES: A widely used digital radiography system based on a photostimulable storage phosphor (PSP) detector was analyzed with regard to radiographic contrast changes that result from the adjustment of detector latitude (x-ray sensitivity range) in the normal processing of chest images. METHODS: Images of an acrylic step wedge were acquired using the digital system in a mode that permitted direct control of effective detector latitude. The images were post-processed in conditions duplicating those used for portable chest examinations, and contrast was measured. RESULTS: Increases in effective detector latitude provided only marginal radiographic contrast gains in the subdiaphragm-equivalent areas of the laser-printed digital film image, while causing large reductions in radiographic contrast in the lung-equivalent region. CONCLUSION: Detector latitude is an important variable that should be monitored or controlled in investigations that compare reader performance using conventional and digital systems.
No takes yet. Share an insight, caveat, or question.
Chotas et al. (1992) studied this question.