Key result
Terminal digit preference and sphygmomanometer calibration errors caused false positive and false negative blood pressure evaluations in 0.82% to 5.18% of the population.
Observational
Systematic errors in blood pressure measurement, such as terminal digit preference and calibration errors, can lead to significant misclassification of hypertension status and inappropriate treatment.
Answers are grounded in this paper and cite their sources.
Measurement errors may prompt inappropriate hypertension treatment; leaves open whether standardized protocols improve classification accuracy.
The accuracy of blood pressure (BP) measurement is important; systematic small errors can mislabel BP status in many persons. The objective of this study was to assess the impact of 2 types of measurement errors on the evaluation of BP in family medicine: errors associated with terminal digit preference and those associated with calibration errors of sphygmomanometers. Secondary data analyses from 2 different projects were used to derive empiric distributions of terminal digit and BP device errors. Taking into account both types of errors, the proportion of false positives (falsely high BP) and false negatives (falsely normal BP) varied between 0. 82% and 5.18% of the population of consulting family physicians. In the United States, false positives and false negatives in patients' BP evaluations might lead to overtreating or undertreating 1.15 million to 7.25 million patients. Results support the need for the development of systematic interventions for quality control of BP measurements and periodic retraining for health professionals.
No takes yet. Share an insight, caveat, or question.
Niyonsenga et al. (2008) conducted an observational in Hypertension. Terminal digit preference and sphygmomanometer calibration errors was evaluated on Proportion of false positives (falsely high BP) and false negatives (falsely normal BP). Terminal digit preference and sphygmomanometer calibration errors caused false positive and false negative blood pressure evaluations in 0.82% to 5.18% of the population.
Synapse has enriched 10 related papers. Consider them for comparative context: