Key result
StentBoost enhanced x-ray imaging showed improved agreement with intravascular ultrasound for assessing minimum stent diameter when using two projection angles and guide catheter calibration (r=0.62).
Why the study?
Does StentBoost enhanced x-ray imaging accurately measure coronary stent deployment compared to intravascular ultrasound in patients undergoing PCI?
Population
33 patients undergoing percutaneous coronary intervention and stent procedures, with 43 stent deployments…
Comparison
StentBoost computer-enhanced x-ray imaging using… vs Intravascular ultrasound and quantitative…
Design
Cohort, Measurements from each modality were made independently without…
Follow-up
Immediate (intraprocedural)
Authors
Loading...
Optimized StentBoost may aid stent assessment in PCI; leaves open whether it can substitute for IVUS.
Observational (n=33)
Single-blind
No
Does StentBoost enhanced x-ray imaging accurately measure coronary stent deployment compared to intravascular ultrasound in patients undergoing PCI?
Effect estimate: r = 0.62
Computer-enhanced x-ray imaging (StentBoost) provides accurate assessment of coronary stent deployment when using two projection angles and guide catheter calibration, offering a less invasive alternative to IVUS.
Davies et al. (2012) conducted an observational in Coronary artery disease requiring percutaneous coronary intervention (n=33). StentBoost enhanced x-ray imaging vs. Intravascular ultrasound (IVUS) was evaluated on Correlation of minimum stent diameter between StentBoost (using two views and guide catheter calibration) and IVUS (r = 0.62). StentBoost enhanced x-ray imaging showed improved agreement with intravascular ultrasound for assessing minimum stent diameter when using two projection angles and guide catheter calibration (r=0.62).