Key result
StentBoost enhanced x-ray imaging showed improved agreement with intravascular ultrasound for assessing minimum stent diameter when using two projection angles and guide catheter calibration.
Why the study?
Does StentBoost enhanced x-ray imaging accurately measure coronary stent deployment compared to intravascular ultrasound in patients undergoing PCI?
Observational (n=33)
Single-blind
No
Does StentBoost enhanced x-ray imaging accurately measure coronary stent deployment compared to intravascular ultrasound in patients undergoing PCI?
Computer-enhanced x-ray imaging (StentBoost) provides accurate assessment of coronary stent deployment when using two projection angles and guide catheter calibration, offering a less invasive alternative to IVUS.
Answers are grounded in this paper and cite their sources.
Optimized StentBoost may aid stent assessment in PCI; leaves open whether it can substitute for IVUS.
OBJECTIVE: To investigate the accuracy of stent measurements using coronary x-ray angiograms with a computer based stent enhancement algorithm applied (StentBoost, SB). To derive recommendations for best practice when using such systems. BACKGROUND: Computer enhancement algorithms allow better visualization of intracoronary stents to assist in ensuring adequate stent deployment. Factors that affect the accuracy of measurements taken on such systems are yet to be fully understood. METHODS: We analysed stent deployment of 43 stents in 33 patients measuring minimum stent diameter and cross sectional area (CSA) using intravascular ultrasound (IVUS), SB enhanced x-ray images, and quantitative coronary angiography (QCA). We investigated if the use of two projections and method of calibration influenced correlation between IVUS and SB measurements. RESULTS: Using two views and performing calibration via the guide catheter improved agreement between SB and IVUS measurements. For example, minimum stent diameter assessed with SB using one view and balloon markers for calibration produced a correlation coefficient, r, of 0.21, whereas using two views and the guide catheter for calibration increased agreement to r = 0.62. Relative measures of stent deployment, such as the ratio of minimum to maximum CSA, produced good correlation between IVUS and SB (r = 0.74). CONCLUSIONS: When using the SB system, two projection angles should be used to image the stent. For absolute measurements, the guide catheter should be used for calibration purposes. Relative measures of stent size, which are probably sufficient for assessment of deployment, also give good agreement with similar measures on IVUS, and require no calibration.
No takes yet. Share an insight, caveat, or question.
Davies et al. (2012) conducted an observational in Coronary artery disease requiring percutaneous coronary intervention (n=33). StentBoost enhanced x-ray imaging vs. Intravascular ultrasound (IVUS) was evaluated on Correlation of minimum stent diameter between StentBoost (using two views and guide catheter calibration) and IVUS (r = 0.62). StentBoost enhanced x-ray imaging showed improved agreement with intravascular ultrasound for assessing minimum stent diameter when using two projection angles and guide catheter calibration.