Key result
Extracorporeal shockwave lithotripsy caused no mechanical damage or detection failure in implantable defibrillators, but post-shock pacing failed in demand mode due to artifact oversensing.
Why the study?
Does exposure to new-generation lithotripter shockwaves affect the mechanical integrity or electrical function of implantable cardioverter defibrillators?
Does exposure to new-generation lithotripter shockwaves affect the mechanical integrity or electrical function of implantable cardioverter defibrillators?
Extracorporeal shockwave lithotripsy does not mechanically damage ICDs or impair detection, but may cause pacing failure via artifact oversensing, prompting a recommendation for R-wave synchronous shockwave application.
PURPOSE: The influence of shockwaves applied during extracorporeal shockwave lithotripsy on the function of implantable cardioverter defibrillators was evaluated. Mechanical influences as well as proper electrical function were tested in an experimental approach. MATERIALS AND METHODS: Two implantable defibrillators (Ventak Mini 1743 and AVII 1821) were exposed to the shockwaves of a new-generation lithotripter. Each of the antidysrhythmic devices was tested at several distances from and within the focus of the lithotripter. All studies were performed with maximum energy and the number of shockwaves used for stone treatment. The devices were connected to an ECG simulator, and continuous recording of a surface ECG, a shock ECG, and marker channel was performed. RESULTS: No macroscopic and microscopic mechanical damage was observed. The detection function of the implantable defibrillators was not altered by any electromagnetic artifacts even when brought into the focus of the shockwaves. All induced ventricular dysrhythmias were terminated properly regarding artifact sensing. However, after defibrillation, the pacing function of the Ventak Mini cardioverter defibrillator, which was programmed into the demand mode, failed. In this case, there was no post-shock pacing in the period of post-defibrillation asystole. The failure was caused by artifact oversensing. CONCLUSION: In patients with implanted cardiac devices undergoing treatment with a new-generation lithotripter, deactivation of the defibrillator is not mandatory. For safety reasons, continuous ECG recording is recommended. To avoid pacing failure by artifact oversensing, the shockwaves should be applied in a R-wave synchronous mode.
No takes yet. Share an insight, caveat, or question.
Küfer et al. (2001) studied Implantable cardioverter defibrillators (n=2). Extracorporeal shockwave lithotripsy (new-generation lithotripter) was evaluated on Mechanical damage and electrical function (detection, dysrhythmia termination, pacing). Extracorporeal shockwave lithotripsy caused no mechanical damage or detection failure in implantable defibrillators, but post-shock pacing failed in demand mode due to artifact oversensing.