Key result
High-frequency scanning acoustic microscopy demonstrated protein adhesion and wall distortion with 0.2 microm spatial resolution on the inner surface of a total artificial heart after implantation.
Why the study?
Can a scanning acoustic microscope evaluate the inner-surface morphology and material fatigue of a total artificial heart?
Can a scanning acoustic microscope evaluate the inner-surface morphology and material fatigue of a total artificial heart?
High-frequency ultrasound using a scanning acoustic microscope is feasible for evaluating material fatigue and protein adhesion in total artificial hearts.
Answers are grounded in this paper and cite their sources.
Provides proof-of-concept in animals; leaves open clinical translation for total artificial heart monitoring.
The total artificial heart (TAH) is being developed for permanent replacement of the natural heart instead of heart transplantation. The need for detecting the material fatigue in the TAH is increasing in order to guarantee long-term use. In this study, the inner surface morphology of the TAH was evaluated by a specially developed scanning acoustic microscope (SAM) system operating in the frequency range of 100-200 MHz. The inner sac of our TAH consisted of polyvinylchloride coated with polyurethane, and the SAM investigations were performed before and after the implantations in goats. The amplitude images of the SAM demonstrated protein adhesion on the inner surface of the TAH after the animal experiment, and the phase images showed distortion of the wall with spatial resolution of 0.2 microm. These results suggest the feasibility of a high-frequency ultrasound for evaluating the material fatigue of TAH.
No takes yet. Share an insight, caveat, or question.
Saijo et al. (2000) studied Total artificial heart material fatigue. Scanning acoustic microscope (SAM) system vs. Before implantation was evaluated on Inner surface morphology (protein adhesion and wall distortion). High-frequency scanning acoustic microscopy demonstrated protein adhesion and wall distortion with 0.2 microm spatial resolution on the inner surface of a total artificial heart after implantation.