Key result
Magnetic resonance imaging accurately and reproducibly measured small right ventricular volumes in fetal lamb heart casts, with 3 mm axial sections producing the best results.
Why the study?
Does MR imaging accurately measure small right ventricular volumes compared to water displacement in a fetal lamb model?
Does MR imaging accurately measure small right ventricular volumes compared to water displacement in a fetal lamb model?
MR imaging can accurately measure small ventricular volumes expected in infants, particularly when using thin (3 mm) axial sections.
Answers are grounded in this paper and cite their sources.
Supports MRI feasibility for sub-milliliter RV volumes in animal models; leaves open human neonatal validation before clinical use.
RATIONALE AND OBJECTIVES: Cine-magnetic resonance (MR) imaging is established as an accurate method of determining cardiac ventricular volumes in adults but has not been validated for the smaller ventricular volumes found in infants. The purpose of the authors study was to evaluate the accuracy and reproducibility of MR imaging to measure small ventricle volumes. METHODS: The volume of silicone-rubber casts of the right ventricle of 11 fetal lamb hearts was measured by MR imaging using different imaging planes and slice section thicknesses and by water displacement. RESULTS: The ventricle volumes of the 11 hearts ranged from 0.45 to 3.00 mL. Accurate and reproducible values for the right ventricle volumes were obtained by MR imaging. Thin sections (3 mm) in the axial plane produced the best results. CONCLUSIONS: Based on studies in the nonmoving heart, MR imaging can accurately measure the small ventricular volumes expected in infants.
No takes yet. Share an insight, caveat, or question.
STAUFFER et al. (1995) studied Small ventricular volumes (n=11). Magnetic Resonance Imaging vs. Water displacement was evaluated on Right ventricle volume measurement accuracy. Magnetic resonance imaging accurately and reproducibly measured small right ventricular volumes in fetal lamb heart casts, with 3 mm axial sections producing the best results.