Key result
The Omron HEM-608 wrist monitor produced readings similar to the standard indirect method, with mean differences of -2.3 ± 7.9 mmHg for systolic and 0.97 ± 5.5 mmHg for diastolic blood pressure.
Why the study?
Does the Omron Model HEM-608 automated wrist monitor provide accurate blood pressure measurements compared to the standard indirect method in normotensive and hypertensive subjects?
Observational (n=100)
Does the Omron Model HEM-608 automated wrist monitor provide accurate blood pressure measurements compared to the standard indirect method in normotensive and hypertensive subjects?
The Omron HEM-608 automated wrist monitor provides blood pressure measurements comparable to standard indirect methods, validating its clinical use.
Answers are grounded in this paper and cite their sources.
Supports wrist BP device accuracy in this cohort; leaves open broader validation before clinical adoption.
OBJECTIVE - The aim of our study was to assess the profile of a wrist monitor, the Omron Model HEM-608, compared with the indirect method for blood pressure measurement. METHODS - Our study population consisted of 100 subjects, 29 being normotensive and 71 being hypertensive. Participants had their blood pressure checked 8 times with alternate techniques, 4 by the indirect method and 4 with the Omron wrist monitor. The validation criteria used to test this device were based on the internationally recognized protocols. RESULTS - Our data showed that the Omron HEM-608 reached a classification B for systolic and A for diastolic blood pressure, according to the one protocol. The mean differences between blood pressure values obtained with each of the methods were -2.3 +7.9mmHg for systolic and 0.97+5.5mmHg for diastolic blood pressure. Therefore, we considered this type of device approved according to the criteria selected. CONCLUSION - Our study leads us to conclude that this wrist monitor is not only easy to use, but also produces results very similar to those obtained by the standard indirect method.
No takes yet. Share an insight, caveat, or question.
Plavnik et al. (2001) conducted an observational in Normotension and hypertension (n=100). Omron Model HEM-608 wrist monitor vs. Standard indirect method was evaluated on Mean differences between blood pressure values obtained with each method (MD -2.3 mmHg (systolic) and 0.97 mmHg (diastolic)). The Omron HEM-608 wrist monitor produced readings similar to the standard indirect method, with mean differences of -2.3 ± 7.9 mmHg for systolic and 0.97 ± 5.5 mmHg for diastolic blood pressure.