Key result
Only an experimental device using 2 von Bahr screws with acrylic cement restored normal femoral stiffness in osteoporotic cadaver femora, while all commercial devices decreased stiffness.
Why the study?
Does the type of osteosynthesis device restore normal femoral stiffness after subcapital osteotomy in osteoporotic bone compared to the intact femur?
Does the type of osteosynthesis device restore normal femoral stiffness after subcapital osteotomy in osteoporotic bone compared to the intact femur?
In an in-vitro model of osteoporotic femoral neck fractures, only a combination of screws and bone cement restored normal femoral stiffness, while all tested commercially available devices failed to do so.
Experimental screw-cement constructs may restore stiffness; leaves open clinical translation in osteoporotic femoral neck fractures.
Restitution of normal bone mechanics after osteosynthesis of a proximal femoral fracture in the elderly is assumed to be a prerequisite for optimal healing. To find the most appropriate fixation device for this purpose, the stiffness of 79 cadaver femora from donors with a history of osteoporosis was measured during axial loading before and after a subcapital osteotomy stabilized with one of nine different osteosynthesis devices. Only an osteotomy stabilized with an experimental device consisting of 2 von Bahr screws (Ericsson AB, Göteborg, Sweden) supported by acrylic cement gave similar mean values as the intact femur. A femur stabilized with the Deyerle device (Biomet, Bridge End, Wales, U.K.), with cannulated screws type Uppsala (Olmed AB, Uppsala, Sweden), or with von Bahr screws had an approximately 20-30% lower stiffness than tested intact. However, these devices provided a higher femoral stiffness than the hook-pin technique type LiH (Söderström AB, Gothenburg, Sweden), cannulated screws (Smith & Nephew, Memphis, TN, U.S.A.), a sliding screw plate (Smith & Nephew), a compression screw with variable length (Biomet), or an experimental screw providing expansion and compression. Despite attempted compression of the osteotomy by some of the devices, all of the commercially available devices resulted in a decrease of femoral stiffness after fixation, which may adversely influence the healing of femoral neck fractures in the elderly. Only a combination of screws and bone cement resulted in normal femoral stiffness, probably because of better device anchorage in the osteoporotic cancellous bone. However, this combination might have the potential risk of vascular damage.
No takes yet. Share an insight, caveat, or question.
Hernefalk et al. (1995) studied Osteoporosis and proximal femoral fracture (n=79). Nine different osteosynthesis devices vs. Intact femur (before osteotomy) was evaluated on Femoral stiffness during axial loading. Only an experimental device using 2 von Bahr screws with acrylic cement restored normal femoral stiffness in osteoporotic cadaver femora, while all commercial devices decreased stiffness.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: