Key result
Simulated defibrillation test protocols yielded probabilities of failing 20-J and 25-J implant criteria ranging from 0.9% to 6.3% and 0.3% to 3.4%, respectively.
Why the study?
How do different defibrillation test protocols influence the probability of meeting implant criteria in patients receiving a defibrillator?
How do different defibrillation test protocols influence the probability of meeting implant criteria in patients receiving a defibrillator?
The choice of defibrillation test protocol significantly impacts the probability of meeting implant criteria, highlighting the need for standardized testing protocols.
Answers are grounded in this paper and cite their sources.
Variable failure rates across protocols warrant caution in testing selection; leaves open optimal standards for defibrillator implantation.
BACKGROUND: Defibrillation testing is a common procedure at defibrillator implant, with the purpose to ensure that each patient receives a device-lead system with a sufficient shock efficacy. The objective of this paper was to study the influence of defibrillation test protocols on the probability of passing implant testing. METHODS: Defibrillation shock efficacy as a function of shock energy was modeled by a dose-response relationship estimated from the clinical data of the PainFREE Rx II study on 564 patients. A Monte Carlo method was used to simulate the outcomes of 12 commonly used defibrillation efficacy test protocols: four safety margin tests and eight protocols estimating the defibrillation threshold (DFT). RESULTS: The probabilities of failing 20-J and 25-J implant criteria for the different protocols ranged from 0.9% to 6.3% for 20 J and 0.3% to 3.4% for 25 J. Large variations in consecutively measured DFT values in the same patients were observed. Best results in the identification of “high risk” patients were obtained with the 2/2 safety margin protocol with an implant criterion of 20 J. The study also showed that the probability of patients inappropriately passing the implant criterion increased when the defibrillation test was repeated after initial failure. CONCLUSION: The defibrillation test protocol greatly influences the probability of meeting implant criterion. Therefore, these test protocols should be standardized. The model developed in this study, could be used to further understand their impact and to derive recommendations.
No takes yet. Share an insight, caveat, or question.
Smits et al. (2011) studied Defibrillator implant (n=564). Defibrillation test protocols was evaluated on Probability of failing 20-J and 25-J implant criteria. Simulated defibrillation test protocols yielded probabilities of failing 20-J and 25-J implant criteria ranging from 0.9% to 6.3% and 0.3% to 3.4%, respectively.