August 1, 1997Pacing and Clinical Electrophysiology

Derivation of a Defibrillator Implant Criterion Based on Probability of Successful Defibrillation

View Full Paper
Ask AI
Bookmark
Share

Key result

A statistical model based on defibrillation efficacy curves from over 1,500 implants was derived to establish an implant criterion meeting predetermined targets for successful defibrillation.

Population

Data from over 1,500 defibrillator implants, including successful epicardial and transvenous electrode…

Design

Other

Authors

PDPaul DeGrootMedtronic (Ireland)TCTimothy R. ChurchUniversity of MinnesotaRMRahul MehraBirla Institute of Technology and Science - Hyderabad Campus

Discussion

Loading...

Member takes

Implication

May support modeling-based defibrillator implant criteria; leaves open prospective validation before clinical adoption.

Structured PICO

P
Population
Data from over 1,500 defibrillator implants used to derive a statistical model for implant criteria based on defibrillation efficacy.
E
Exposure
Derivation of a defibrillator implant criterion using a statistical model based on defibrillation efficacy curves
O
Outcome
Probability of successful defibrillation by a single shock at a predetermined maximum output

A statistical model based on defibrillation efficacy curves can be used to establish evidence-based defibrillator implant criteria, potentially replacing empirical safety margins.

Cite This Study

DeGroot et al. (1997) studied Cardioverter defibrillator implantation (n=1,500). Statistically derived defibrillator implant criterion vs. Empirical criteria (2:1 energy safety margin or fixed 10 joules) was evaluated on Probability of successful defibrillation. A statistical model based on defibrillation efficacy curves from over 1,500 implants was derived to establish an implant criterion meeting predetermined targets for successful defibrillation.

synapsesocial.com/papers/6aa380ed9fb7966fc9748e36https://doi.org/10.1111/j.1540-8159.1997.tb03598.x

Topics

Cardiac device therapy
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1The Defibrillation Success Rate Versus Energy Relationship: Part I‐Curve Fitting and the Most Efficient Defibrillation Energy1990 · 42 citations
  2. 2A Model to Evaluate Alternative Methods of Defibrillation Threshold Determination1988 · 28 citations
  3. 3The relationship between successful defibrillation and delivered energy in open-chest dogs: Reappraisal of the “defibrillation threshold” concept1987 · 217 citations
  4. 4Measurement of cardioversion/defibrillation thresholds in man by a truncated exponential waveform and an apical patch-superior vena caval spring electrode configuration.1984 · 109 citations
  5. 5Estimating the 95% effective defibrillation dose1993 · 30 citations