This is the authors' abstract. We don't add key points for this paper.
A wireless inductive sensing technology with a robust packaging scheme is proposed and demonstrated for future implantable hydrogel-based glucose monitoring applications. A prototype wireless glucose sensor employs a glucose-sensitive hydrogel embedding a parylene-coated metallic sensing plate as a key sensing element. The hydrogel changes its volume responding to glucose change; thus displacing the metallic sensing plate that can be sensed by a nearby sensing coil through the coil inductance change. An ultra-low-power tunnel diode oscillator is designed to convert the inductance change to an oscillator frequency change, which can be wirelessly detected by a nearby receiver. The entire sensing system can be energized by RF powering, thus achieving a wireless and battery-less operation. The proposed sensing scheme allows the sensing electronics to be fully encapsulated in a hermetically sealed package, critical for long-term implantable applications. In vitro characterization demonstrates that the prototype sensor operated around 40 MHz can be wirelessly powered by a 2.77 mW 5 MHz power source with a distance of 2 cm, and achieves a sensing resolution of 0.05 mM and 0.3 mM within a glucose range from 0 mM to 2 mM and 8 mM to 10 mM, respectively, limited by the oscillator frequency drift and fluctuation. Furthermore, a hydrogel pillar-based design is promising to substantially shorten the system response time.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Yu et al. (2019) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: