Key result
Five cases of longitudinal shortening with the Promus Element stent were observed over 15 months, supporting bench-top data correlating its two-link design with compression risk.
Why the study?
Does the Promus Element stent platform have a risk of longitudinal shortening?
Case Report (n=5)
No
Does the Promus Element stent platform have a risk of longitudinal shortening?
This case series highlights a potential safety concern regarding longitudinal shortening and compression risk with the Promus Element stent platform.
Hypothesis-generating for longitudinal shortening with Promus Element; larger studies needed to quantify incidence and clinical impact.
This case series raises further awareness of the potential for longitudinal shortening with the Promus Element™ (Boston Scientific, MN) drug-eluting stent (DES) platform. With safety at the forefront of DES development, newer generation systems are utilizing thinner platforms with little or no polymer. Although these contemporary, ultrathin platforms also improve conformability and deliverability, specific design characteristics might inadvertently reduce longitudinal strength. To date, there are no randomized data addressing this issue in vivo for any stent platform, with opinion based on individual experience and bench testing. We report five cases of longitudinal shortening with the Promus Element stent occurring in our institution over a 15-month period. During the same period, we experienced only isolated cases with other second generation DES platforms. Our experience supports recently published bench-top data correlating the two-link design of this platform with the risk of compression.
No takes yet. Share an insight, caveat, or question.
Shannon et al. (2012) conducted a case report in Patients receiving drug-eluting stents (n=5). Promus Element drug-eluting stent vs. Other second generation DES platforms was evaluated on Longitudinal shortening. Five cases of longitudinal shortening with the Promus Element stent were observed over 15 months, supporting bench-top data correlating its two-link design with compression risk.