Key result
Distal radial arterial pressure monitoring showed good agreement with forearm radial arterial pressure during prone spinal surgery.
Why the study?
Agreement and interchangeability between distal and forearm radial arterial pressures during general anesthesia for prone spinal surgery needed to be tested.
Does distal radial arterial pressure monitoring provide comparable measurements to forearm radial arterial pressure monitoring in patients undergoing prone spinal surgery?
Observational (n=40)
Does distal radial arterial pressure monitoring provide comparable measurements to forearm radial arterial pressure monitoring in patients undergoing prone spinal surgery?
Distal radial arterial pressure monitoring shows good agreement with forearm radial arterial pressure monitoring and may be a suitable substitute during prone spinal surgery.
Answers are grounded in this paper and cite their sources.
May support distal radial monitoring as alternative in prone spinal surgery; leaves open need for randomized validation.
OBJECTIVE: To test agreement and interchangeability between distal (dRA) and forearm radial arterial (RA) pressures (AP) during general anesthesia (GA) for prone spinal surgery. METHODS: This prospective observational study involved 40 patients scheduled for GA spinal surgery. The right dRA and left forearm RA were cannulated in all patients to continuously measure invasive blood pressures (IBP). We compared the agreement and trending ability of systolic AP (SAP), diastolic AP (DAP), and mean AP (MAP) at each site 15 minutes after tracheal intubation, start of surgery, 30 and 60 minutes after the start of surgery, and after skin suturing. RESULTS: Paired BP values (n = 184) (37 cases) were analyzed. The bias (standard deviation), limits of agreement, and percentage error were: SAP: 0.19 (3.03), -5.75 to 6.12, and 5.04%; DAP: -0.06 (1.75), -3.50 to 3.38, and 5.10%; and MAP: 0.08 (1.52), -2.90 to 3.05, and 3.54%, respectively. The linear regression coefficients of determination were 0.981, 0.982, and 0.988 for SAPs, DAPs, and MAPs, respectively; four-quadrant plot concordance rates were 95.11%, 92.03%, and 92.66%, respectively. CONCLUSION: All arterial BPs showed good agreement and trending capabilities for both the dRA and RA. The dRA may be substituted for the RA in IBP monitoring.
No takes yet. Share an insight, caveat, or question.
Xu et al. (2023) conducted an observational in Prone spinal surgery (n=40). Distal radial arterial (dRA) pressure monitoring vs. Forearm radial arterial (RA) pressure monitoring was evaluated on Agreement and trending ability of systolic, diastolic, and mean arterial pressures. Distal radial arterial pressure monitoring showed good agreement with forearm radial arterial pressure during prone spinal surgery.