Key result
A single lead Lumax 640 DX ICD detected an unusual pattern of artifacts related to electromagnetic interference that were misclassified as atrial tachycardia.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
The novel DX ICD system can detect electromagnetic interference as high-frequency episodes on far-field and atrial channels, potentially leading to misclassification as atrial tachycardia.
A 56-year-old man underwent a single lead Lumax 640 DX implantable cardioverter defibrillator implantation for primary prevention of sudden cardiac death. A DX system consists of a single lead, which provides atrial as well as ventricular electrograms, and enhances atrial arrhythmia detection. Three months after the implantation, high-frequency episodes were detected on the far field and the atrial channels, but not on the bipolar right ventricular channel; these were classified as atrial tachycardia. In the present report, we discussed the unusual pattern of the artifacts that was related to an electromagnetic interference detected by the novel DX system.
No takes yet. Share an insight, caveat, or question.
Konstantino et al. (2016) conducted a case report in Primary prevention of sudden cardiac death (n=1). Single lead Lumax 640 DX implantable cardioverter defibrillator was evaluated on Detection of artifacts related to electromagnetic interference. A single lead Lumax 640 DX ICD detected an unusual pattern of artifacts related to electromagnetic interference that were misclassified as atrial tachycardia.