Key result
The new software 'SLDF version 4.0' demonstrated high reliability for assessing retinal arteriolar structural parameters, with coefficients of variation significantly less than 10% for most measures.
Why the study?
Does the new software 'SLDF version 4.0' improve the reliability of measuring retinal arteriolar morphology compared to the old software in healthy persons and hypertensive patients?
Observational (n=28)
Does the new software 'SLDF version 4.0' improve the reliability of measuring retinal arteriolar morphology compared to the old software in healthy persons and hypertensive patients?
The new SLDF version 4.0 software provides highly reliable in vivo measurements of retinal arteriolar structure, which is useful for assessing arteriolar remodeling in hypertension.
Answers are grounded in this paper and cite their sources.
Supports reliable retinal arteriolar assessment in hypertension research; leaves open clinical adoption pending prospective validation.
OBJECTIVE: The investigation of the retinal arterioles offers the unique opportunity to analyze in vivo arteriolar remodeling in arterial hypertension in humans. We analyzed the reliability of assessing retinal arteriolar morphology with our new version of the software analyses for scanning laser Doppler flowmetry. METHOD: In the test-retest reliability study, 10 eyes of 10 healthy persons were measured during 5 days under routine laboratory conditions with the Heidelberg Retinal Flowmetry. In a second study, interobserver and intraobserver reliability was analyzed from retinal images of 18 patients with three types of arterial hypertension by three readers and the most experienced reader analyzed all images twice on two different days. Images were analyzed by the old and the newly developed software versions. To characterize the reliability, the coefficients of variation were calculated. RESULTS: The test-retest study analyzed with the new program showed that the variation coefficients of vessel and lumen diameter, wall thickness, wall/lumen ratio and new calculated parameter: lumen/vessel diameter ratio of retinal arterioles were significantly less than 10%, with the exception of the wall cross-sectional area (12.5%). The interobserver and intraobserver reliability showed in nearly all circumstances coefficients of variations of less than 10% and did not differ across various readers and patient groups. CONCLUSION: The new software 'SLDF version 4.0' clearly improved the reliability of assessing the structural parameters of the retinal arterioles. The application delivers reliable measurements of the retinal arteriolar structure in vivo in humans.
No takes yet. Share an insight, caveat, or question.
Haraźny et al. (2011) conducted an observational in Arterial hypertension (n=28). New software 'SLDF version 4.0' for scanning laser Doppler flowmetry vs. Old software version was evaluated on Coefficients of variation of retinal arteriolar structural parameters. The new software 'SLDF version 4.0' demonstrated high reliability for assessing retinal arteriolar structural parameters, with coefficients of variation significantly less than 10% for most measures.