Key result
The total balloon inflation volume for the IntraClude device was strongly correlated with ascending aorta diameter, yielding the formula: volume = 1.54 × diameter - 20.81.
Why the study?
To evaluate an IntraClude balloon filling strategy based on the relationship between balloon pressure, volume, and ascending aorta size instead of manufacturer-suggested incremental filling.
Observational (n=33)
A linear relationship between ascending aorta diameter and required saline volume can predict the optimal inflation volume for the IntraClude endo-balloon during minimally invasive mitral valve surgery.
Answers are grounded in this paper and cite their sources.
May inform IntraClude balloon volumes in minimally invasive mitral surgery; leaves open prospective validation before clinical adoption.
BACKGROUND: IntraClude® Edwards, the endo-balloon for aortic cross-clamping in heart-port procedures, is featured by a new design, resulting in an easier clamping technique. The producer suggests a balloon filling strategy based on an initial volume of 10 mL of saline, followed by additional injections of 2 mL up to complete aortic occlusion. We report a filling procedure based on a relationship between balloon pressure and volume and the size of ascending aorta.METHODS: From August 2012 to November 2015, 33 patients underwent minimally invasive mitral valve surgery with IntraClude® aid. During the first four procedures, aortic occlusion appeared complete when the balloon pressure was kept above 330 mmHg; an additional 2 mL volume was injected whenever this value got closed to 300 mmHg. We analyzed the final balloon volume and the aortic diameter of each patient within a balloon pressure between 300 and 350 mmHg.RESULTS: Mean ascending aorta diameter was 31.45±2.12 mm, mean cross-clamping time 120.73±41.01 minutes, mean saline volume inflated 27.73±4.24 mL, mean balloon pressures 340.93±53.74 mmHg. The relationship between ascending aorta diameter and total balloon inflation volume calculated by linear regression was: inflation volume=1.54×ascending aorta diameter -20.81. Statistical analysis showed an R=0.84 (P<0.001), R2=0.7 with an adjusted R2=0.69 and a standard error of estimate=4.06.CONCLUSIONS: A relationship correlating ascending aorta dimensions and saline volume necessary to reach an effective aortic cross clamp is useful to guarantee a stable occlusion throughout surgery. Additional injections can be required to keep the balloon pressure between the 300 and 350 mmHg.
No takes yet. Share an insight, caveat, or question.
Contino et al. (2021) conducted an observational in Minimally invasive mitral valve surgery (n=33). IntraClude device was evaluated on Relationship between ascending aorta diameter and total balloon inflation volume (R=0.84, p=<0.001). The total balloon inflation volume for the IntraClude device was strongly correlated with ascending aorta diameter, yielding the formula: volume = 1.54 × diameter - 20.81.