Key result
Valve-in-valve TAVI using the 20-mm Edwards Sapien 3 transcatheter heart valve in 5 patients with degenerated 19 mm bioprosthetic valves was technically feasible with no mild aortic insufficiency.
Why the study?
Is valve-in-valve TAVI using a 20-mm Edwards Sapien 3 valve feasible and effective in patients with degenerated 19 mm bioprosthetic aortic valves?
Case Report (n=5)
Is valve-in-valve TAVI using a 20-mm Edwards Sapien 3 valve feasible and effective in patients with degenerated 19 mm bioprosthetic aortic valves?
Valve-in-valve TAVI using a 20-mm Edwards Sapien 3 valve is technically feasible for degenerated 19 mm bioprosthetic aortic valves, providing good hemodynamics without significant paravalvular leak.
Small series supports feasibility of S3 valve-in-valve TAVI in degenerated 19-mm valves; leaves open durability and broader safety.
INTRODUCTION: Transcatheter aortic valve implantation (TAVI) in patients with degenerated bioprosthetic aortic valve has been successfully performed as an alternative to surgery. We describe our initial experience of valve-in-valve TAVI in five patients, using new generation Edwards Sapien 3 transcatheter heart valves implanted into degenerated 19 mm bioprosthetic valves. 20-mm Edwards S3 valves were offered for compassionate use. All patients had significant aortic valve stenosis. METHODS AND RESULTS: The main vascular access was achieved and pre-closed with two Proglide closure devices in one patient and Prostar closure devices in four patients. For each TAVI procedure an Edwards 14 French sheath was inserted without complication and sutured in place. The Sapien 3 Commander delivery system was inserted and the valve was aligned in the descending aorta. The 20-mm Sapien 3 valve was deployed with slow continuous inflation during rapid right ventricular pacing. The cranial edge of the Edwards S3 valve was aligned with the cranial radiopaque markers of bioprosthesis to minimize paravalvular leak. Post-deployment angiography, transesophageal echocardiography and aortogram confirmed absence of mild aortic insufficiency and no increase in trans-aortic gradient when compared to a naïve 19 mm bioprosthetic valve. CONCLUSION: Valve-in-valve TAVI with the Edwards S3 transcatheter heart valve for degenerative bioprosthetic aortic valves is technically feasible. The proper position of the stented valve minimizes the risk for post-procedure paravalvular insufficiency and provides good transaortic pressure gradient.
No takes yet. Share an insight, caveat, or question.
Loyalka et al. (2016) conducted a case report in Degenerated 19 mm bioprosthetic aortic valve with significant aortic valve stenosis (n=5). Valve-in-valve TAVI with 20-mm Edwards Sapien 3 transcatheter heart valve was evaluated on Technical feasibility, paravalvular leak, and trans-aortic gradient. Valve-in-valve TAVI using the 20-mm Edwards Sapien 3 transcatheter heart valve in 5 patients with degenerated 19 mm bioprosthetic valves was technically feasible with no mild aortic insufficiency.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: