Key result
A flow estimation method for an implantable centrifugal blood pump was evaluated across 79 in vitro operating points and during a 40-day in vivo goat experiment.
A novel flow estimation method for an implantable centrifugal blood pump was developed and tested in vitro and in vivo, offering a potentially less invasive and non-thrombogenic monitoring approach.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for sensorless flow monitoring in centrifugal pumps; human validation required before clinical consideration.
A less invasive and non thrombogenic flow estimation of an implantable centrifugal blood pump (ICBP) has been developed, which was derived from electric power consumption, the rotating speed of a motor, and blood viscosity presumed by hematocrit and body temperature. The power consumption and the rotating speed of the motor were measured by a wattmeter every 0.2 sec. Accuracy and stability of the estimated flow (EF) were investigated during in vitro and in vivo experiments. The EF was compared with a measured flow rate (MF) monitored by an electromagnetic flowmeter. During in vitro experiments, the EF and MF were measured at 79 operating points. The ICBP was driven in a closed mock loop filled with goat blood with hematocrit values of 21.5, 28, 34, and 42%. During in vivo experiments, the ICBP was implanted in the chest cavity of a goat and driven for 40 days
No takes yet. Share an insight, caveat, or question.
Wakisaka et al. (1997) studied this question. Flow estimation of an implantable centrifugal blood pump vs. Measured flow rate by electromagnetic flowmeter was evaluated on Accuracy and stability of the estimated flow. A flow estimation method for an implantable centrifugal blood pump was evaluated across 79 in vitro operating points and during a 40-day in vivo goat experiment.
Synapse has enriched 10 related papers. Consider them for comparative context: