Key result
In a canine model of left ventricular failure, the direct mechanical left ventricular assist device increased cardiac output by 37% and decreased mean left atrial pressure by 15%.
Why the study?
Does a direct mechanical left ventricular assist device improve hemodynamic parameters in a canine model of left ventricular failure?
Does a direct mechanical left ventricular assist device improve hemodynamic parameters in a canine model of left ventricular failure?
A novel direct mechanical left ventricular assist device improved hemodynamic parameters in a canine model of severe left ventricular failure, suggesting potential as a future bridge to transplantation.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for direct mechanical LVADs in LV failure; should not change practice pending human trials.
We have developed a direct mechanical left ventricular assist device (DMLVAD) for severe left ventricular failure. The DMLVAD was attached to the left ventricle and compressed the heart by a pneumatic driving unit. In a mock circulation model with an extracted nonbeating heart, a cardiac output (CO) of 1.93 L/min was obtained at a driving pressure of 200 mm Hg. In a canine left ventricular failure model induced by injection of sodium hydroxide into the myocardium, the systolic arterial pressure, systolic left ventricular pressure, maximum LV dP/dt, peak flow, and CO increased by 21, 24, 58, 144, and 37%, respectively. The mean left atrial pressure also decreased by 15% when the DMLVAD was driven. These effects were most prominent when the mean left atrial pressure was over 15 mm Hg, and the driving pressure was over 100 mm Hg. Compression at late systole was more effective in obtaining greater CO. We suggest that the DMLVAD could be an optional circulatory assist device for patients with left ventricular failure awaiting heart transplantation.
No takes yet. Share an insight, caveat, or question.
HOTEI et al. (1997) studied severe left ventricular failure. Direct mechanical left ventricular assist device (DMLVAD) was evaluated on Hemodynamic parameters including cardiac output and left atrial pressure. In a canine model of left ventricular failure, the direct mechanical left ventricular assist device increased cardiac output by 37% and decreased mean left atrial pressure by 15%.