Key result
A magnetically actuated left ventricular assist device increased mean aortic pressure by 45-50 mmHg and provided pump flow up to 4.6 L/min in an animal model of extreme ventricular dysfunction.
Why the study?
Does a magnetically actuated LVAD maintain satisfactory hemodynamic parameters in an animal model of extreme left ventricular dysfunction?
Does a magnetically actuated LVAD maintain satisfactory hemodynamic parameters in an animal model of extreme left ventricular dysfunction?
A novel magnetically actuated LVAD demonstrated the ability to maintain satisfactory hemodynamics in a sheep model of acute extreme left ventricular dysfunction.
Answers are grounded in this paper and cite their sources.
May inform magnetically actuated LVAD development; leaves open human translation pending clinical trials.
Over the past 6 years, research has led to development of a small, lightweight, power-efficient, uniquely simple ventricular assist device driven by a magnetic actuator. Magnetic actuation permits total elimination of all mechanical motion converter components used for pusher plate displacement, resulting in a significant reduction in complexity and resultant increase in reliability. Extensive in vitro mock loop development has resulted in a left ventricular assist device (LVAD), the primary design parameters of which for the clinical prototype actuator and pump are 1) an actuator weight of 312 g, 2) actuator size of 32.5 cm3, 3) power requirements of 7.8 to 11.4 watts (60-100 beats per minute [BPM]), and 4) system efficiency of 24% to 34% and average dynamic stroke volume of 65 ml. Initial in vivo tests assessed this LVAD's performance in four sheep under three acute conditions of ventricular dysfunction. The results demonstrate that, at a pump-rate of 100 BPM, mean aortic pressure increased by 45-50 mmHg during 1) beta blockade, 2) coronary ligation, and 3) ventricular fibrillation. Pump flow ranged from 2.71 L/min to a maximum of 4.6 L/min. Acute test periods were arbitrarily set for 6 hours duration. Of the four sheep, two survived, one lived 5 hours, and the fourth lived 4.5 hours. Global fibrillation was the primary cause of failure. This initial in vivo data demonstrates the pump's ability to maintain satisfactory hemodynamic parameters of flow and pressure under three acute conditions of extreme left ventricular dysfunction in an animal model. These initial LVAD performances were encouraging. Further tests will use calves with a greatly expanded performance evaluation protocol.
No takes yet. Share an insight, caveat, or question.
Kovacs et al. (1992) studied ventricular dysfunction (n=4). Magnetically actuated left ventricular assist device was evaluated on Hemodynamic parameters of flow and pressure. A magnetically actuated left ventricular assist device increased mean aortic pressure by 45-50 mmHg and provided pump flow up to 4.6 L/min in an animal model of extreme ventricular dysfunction.
Synapse has enriched 10 related papers. Consider them for comparative context: