Key result
Placement of the pulmonary artery wedge catheter above the left atrium with 5 cm H2O of PEEP caused a 5 mm Hg gradient between PAWP and LAP in normovolemic dogs.
Answers are grounded in this paper and cite their sources.
Accurate estimation of left atrial pressure using pulmonary artery wedge pressure requires fluid continuity, which can be disrupted by catheter placement above the left atrium, increased PEEP, or hypovolemia.
Pulmonary artery wedge pressure (PAWP) will only reflect left atrial pressure (LAP) if continuity of fluid exists from the catheter tip to the left atrium. Either increased airway pressure or decreased hydrostatic pressure may lead to discontinuity of the fluid column and midinterpretation of PAWP. Simultaneous measurements of PAWP and LAP were made in 19 anesthetized dogs. Placement of the pulmonary artery wedge catheter above the left atrium (West Zone I) in combination with the incremental addition of 5 cm H2O of PEEP caused a 5 mm Hg gradient between PAWP and LAP in the normovolemic animal. Augmenting PEEP further or hypovolemia (i.e., decrease in LAP) increased the gradient. Hypervolemia (increase in LAP) diminished the gradient. Fluid continuity between the PAW catheter and LA is a prerequisite for monitoring LAP with the Swan-Ganz catheter. Increases in PEEP, placement of the catheter above the left atrium and hypovolemia may occlude the fluid column and cause artifacts in the PAWP obtained.
No takes yet. Share an insight, caveat, or question.
Kane et al. (1978) studied this question. Placement of pulmonary artery wedge catheter above the left atrium with PEEP was evaluated on Gradient between pulmonary artery wedge pressure (PAWP) and left atrial pressure (LAP). Placement of the pulmonary artery wedge catheter above the left atrium with 5 cm H2O of PEEP caused a 5 mm Hg gradient between PAWP and LAP in normovolemic dogs.
Synapse has enriched 10 related papers. Consider them for comparative context: