Key result
Closer bipolar electrode spacing (2 mm) improved detection of local abnormal ventricular activities and increased the near-/far-field voltage ratio compared to 4 mm and 8 mm spacing (p<0.0001).
Why the study?
Does closer bipolar electrode spacing improve the detection of local abnormal ventricular activities and near-field electrograms in infarcted sheep?
Does closer bipolar electrode spacing improve the detection of local abnormal ventricular activities and near-field electrograms in infarcted sheep?
p-value: p=<0.0001
Closer bipolar electrode spacing (2 mm) improves the detection of near-field electrograms and local abnormal ventricular activities by reducing far-field effects in infarcted tissue.
May improve near-field detection in infarcted animal tissue; leaves open whether 2-mm spacing benefits human ventricular tachycardia ablation.
OBJECTIVES: This study sought to evaluate the relation between bipolar electrode spacing and far- and near-field electrograms. BACKGROUND: The detailed effects of bipolar spacing on electrograms (EGMs) is not well described. METHODS: With a HD-Grid catheter, EGMs from different bipole pairs could be created in each acquisition. This study analyzed the effect of bipolar spacing on EGMs in 7 infarcted sheep. A segment was defined as a 2-mm center-to-center bipole. In total, 4,768 segments (2,020 healthy, 1,542 scar, and 1,206 in border areas, as defined by magnetic resonance imaging [MRI]) were covered with an electrode pair of spacing of 2 mm (Bi-2), 4 mm (Bi-4), and 8 mm (Bi-8). RESULTS: A total of 3,591 segments in Bi-2 were free from local abnormal ventricular activities (LAVAs); 1,630 segments were within the MRI-defined scar and/or border area. Among them, 172 (10.6%) segments in Bi-4 and 219 (13.4%) segments in Bi-8 showed LAVAs. In contrast, LAVAs were identified in 1,177 segments in Bi-2; 1,118 segments were within the MRI-defined scar and/or border area. Among them, LAVAs were missed in 161 (14.4%) segments in Bi-4 and in 409 (36.6%) segments in Bi-8. In segments with LAVAs, median far-field voltage increased from 0.09 mV (25th to 75th percentile: 0.06 to 0.14 mV) in Bi-2, to 0.16 mV (25th to 75th percentile: 0.10 to 0.24 mV) in Bi-4, and to 0.28 mV (25th to 75th percentile: 0.20 to 0.42 mV) in Bi-8 (p < 0.0001). Median near-field voltage increased from 0.14 mV (25th to 75th percentile: 0.08 to 0.25 mV) in Bi-2, to 0.21 mV (25th to 75th percentile: 0.12 to 0.35 mV) in Bi-4, and to 0.32 mV (25th to 75th percentile: 0.17 to 0.48 mV) in Bi-8 (p < 0.0001). The median near-/far-field voltage ratio decreased from 1.67 in Bi-2, to 1.43 in Bi-4, and 1.23 in Bi-8 (p < 0.0001). CONCLUSIONS: Closer spacing better discriminates surviving tissue from dead scar area. Although far-field voltage systematically increases with spacing, near-field voltages were more variable, depending on local surviving muscular bundles. Near-field EGMs are more easily observed with smaller spacing, largely due to the reduction of the far-field effect.
No takes yet. Share an insight, caveat, or question.
Takigawa et al. (2018) studied Myocardial infarction (infarcted sheep) (n=7). Bipolar electrode spacing (2 mm, 4 mm, 8 mm) vs. Comparison between 2 mm, 4 mm, and 8 mm spacing was evaluated on Near-/far-field voltage ratio (p=<0.0001). Closer bipolar electrode spacing (2 mm) improved detection of local abnormal ventricular activities and increased the near-/far-field voltage ratio compared to 4 mm and 8 mm spacing (p<0.0001).