Key result
A miniature conductance catheter reliably monitored right ventricular volume changes in rabbit hearts, with measured conductance highly correlated with reference volumes.
Answers are grounded in this paper and cite their sources.
A miniature conductance catheter can reliably monitor in vivo relative right ventricular volume changes in small-animal hearts.
To assess the reliability of conductance (G) catheter for evaluating right ventricular (RV) volume changes, a miniature (3.5F) six-electrode catheter was developed and tested in 11 New Zealand rabbit hearts. In five animals the heart was excised; in six it was left in the thorax. RV conductance was recorded while the RV was filled with blood in 0.25-ml steps at different left ventricular (LV) volumes. Linear correlation of measured conductance vs. reference volumes was computed. RV conductance was highly correlated with reference volume [correlation coefficient (r) ranging from 0.991 to 0.999]. Slope of regression lines was not significantly affected by LV volume variations in 1-ml steps or by acute conductance changes of structures surrounding the heart, whereas the intercept was affected only by the 0- to 1-ml LV volume change. In four rabbits, RV conductance changes during a cardiac cycle [stroke volume- (SV) G] were compared in vivo with electromagnetic flow probe-derived estimates of SV (SVem) as stroke volume was varied by graded inferior vena caval occlusion. SV-G correlated well with SVem (r ranging from 0.92 to 0.96). This correlation persisted after the thorax was filled with saline; however, significant differences were found in individual slopes (P < 0.001). These results show that the conductance catheter has a potential to reliably monitor in vivo relative RV volume changes in small-animal hearts.
No takes yet. Share an insight, caveat, or question.
Soldà et al. (1992) studied Right ventricular volume changes assessment (n=11). Miniature (3.5F) six-electrode conductance catheter vs. Reference volumes and electromagnetic flow probe-derived estimates was evaluated on Correlation of measured conductance vs. reference volumes (r=0.991 to 0.999). A miniature conductance catheter reliably monitored right ventricular volume changes in rabbit hearts, with measured conductance highly correlated with reference volumes.
Synapse has enriched 10 related papers. Consider them for comparative context: