Key result
ECG-recording direction significantly impacts VF-waveform analysis, with AMSA-values in lead I being on average 2.24 lower than in lead II (8.0 vs 10.2; P<0.001).
Why the study?
Given variability in in-field pad placement and new resuscitation protocols restricting prompt defibrillation to high AMSA cases, the impact of ECG recording direction on AMSA values and defibrillation advice was evaluated.
Does ECG recording direction impact ventricular fibrillation waveform measurements (AMSA) in ICD-recipients?
Observational (n=243)
Does ECG recording direction impact ventricular fibrillation waveform measurements (AMSA) in ICD-recipients?
Mean Difference: -2.24
Absolute Event Rate: 8% vs 10.2%
p-value: p=<0.001
ECG-recording direction significantly affects VF-waveform analysis, which may impact shock guidance by smart defibrillators.
ECG lead selection may influence AMSA-guided defibrillation in ICD patients; leaves open whether standardization improves VF outcomes.
AIM: In cardiac arrest, ventricular fibrillation (VF) waveform analysis has identified the amplitude spectrum area (AMSA) as a key predictor of defibrillation success and favorable neurologic survival. New resuscitation protocols are under investigation, where prompt defibrillation is restricted to cases with a high AMSA. Appreciating the variability of in-field pad placement, we aimed to assess the impact of recording direction on AMSA-values, and the inherent defibrillation advice. METHODS: Prospective VF-waveform study on 12-lead surface electrocardiograms (ECGs) obtained during defibrillation testing in ICD-recipients (2010-2017). AMSA-values (mVHz) of simultaneous VF-recordings were calculated and compared between all limb leads, with lead II as reference (proxy for in-field pad position). AMSA-differences between leads I and II were quantified using Bland-Altman analysis. Moreover, we investigated differences between these adjacent leads regarding classification into high (≥15.5), intermediate (6.5-15.5) or low (≤6.5) AMSA-values. RESULTS: In this cohort (n = 243), AMSA-values in lead II (10.2 ± 4.8) differed significantly from the other limb leads (I: 8.0 ± 3.4; III: 12.9 ± 5.6, both p < 0.001). The AMSA-value in lead I was, on average, 2.24 ± 4.3 lower than in lead II. Of the subjects with high AMSA-values in lead II, only 15% were classified as high if based on assessments of lead I. For intermediate and low AMSA-values, concordances were 66% and 72% respectively. CONCLUSIONS: ECG-recording direction markedly affects the result of VF-waveform analysis, with 20-30% lower AMSA-values in lead I than in lead II. Our data suggest that electrode positioning may significantly impact shock guidance by 'smart defibrillators', especially affecting the advice for prompt defibrillation.
No takes yet. Share an insight, caveat, or question.
Thannhauser et al. (2021) conducted an observational in Ventricular fibrillation (n=243). ECG-recording direction (lead I) vs. Lead II (reference) was evaluated on Amplitude spectrum area (AMSA) values (MD -2.24, p=<0.001). ECG-recording direction significantly impacts VF-waveform analysis, with AMSA-values in lead I being on average 2.24 lower than in lead II (8.0 vs 10.2; P<0.001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: