Key result
Intraoperative MediStim transit time flow measurement (TTFM) is highly useful during CABG, as a Pulsatility Index greater than 5 generally indicates a correctable graft problem in off-pump cases.
Answers are grounded in this paper and cite their sources.
Intraoperative transit time flow measurement is highlighted as a useful tool in CABG to identify and correct graft problems when the Pulsatility Index exceeds 5.
This is an important paper as it underscores the importance of utilizing MediStim transit time flow measurement (TTFM) intraoperatively [1]. In my experience, I generally revise all grafts with a Pulsatility Index (P.I.) greater than 5. In off-pump cases, I have found that those grafts with a P.I. above 5 almost always have a correctable problem. The only exceptions have been in those instances where there is a large discrepancy between the size of the conduit and the target vessel and where the target vessel is very short. In on-pump coronary bypass procedures, another important consideration is the existence of spasm resulting from cold areas of the heart following cardioplegia.
No takes yet. Share an insight, caveat, or question.
F. Benetti (2008) conducted a letter in Coronary artery bypass grafting (CABG). MediStim transit time flow measurement (TTFM) was evaluated. Intraoperative MediStim transit time flow measurement (TTFM) is highly useful during CABG, as a Pulsatility Index greater than 5 generally indicates a correctable graft problem in off-pump cases.
Synapse has enriched 10 related papers. Consider them for comparative context: