Key result
Non-Invasive Cardiac System (NICaS) demonstrated a strong positive correlation with continuous thermodilution for measuring cardiac index (r = 0.765), but the parameters are not interchangeable due to wide limits of agreement and a 64% percentage error.
Why the study?
Cardiac output assessment is central to advanced haemodynamic management in critically ill patients, prompting validation of NICaS-derived cardiac index and output against thermodilution in post-operative cardiac surgical patients.
Does continuous cardiac output monitoring derived from regional impedance cardiography accurately correlate with continuous thermodilution in post-operative CABG patients?
Observational (n=23)
No
Does continuous cardiac output monitoring derived from regional impedance cardiography accurately correlate with continuous thermodilution in post-operative CABG patients?
While regional impedance cardiography demonstrates good trending ability for cardiac output, its high percentage error indicates it is not interchangeable with continuous thermodilution in post-operative cardiac surgical patients.
Answers are grounded in this paper and cite their sources.
NICaS shows good trending ability but lacks interchangeability with thermodilution; leaves open its role as a noninvasive alternative pending further validation.
Background: Cardiac output (CO) assessment is a corner stone in advanced haemodynamic management, especially in critical ill patients. The present study was conducted to validate cardiac index and cardiac output by NICaS™ with the thermodilution technique using pulmonary artery catheter in post-operative cardiac surgical patients. Materials and Methods: This was a prospective observational clinical study conducted at a tertiary care hospital. 23 adult patients in the age range of 18-65 years who had undergone for elective coronary artery bypass grafting were included in the study. Results: Spearman's correlation coefficient of cardiac index between continuous Thermodilution (cTD) and Non-Invasive Cardiac System (NICaS™) showed a good correlation (r = 0.765, 95% confidence interval 0.70 to 0.82, P < 0.0001). There was a good correlation between cTD and NICaS™ for cardiac output (r = 0.759, 95% confidence interval 0.69 to 0.81, P < 0.0001), Bland-Altman plot for cardiac index between cTD and NICaS™ showed a mean bias of -0.66 ± 0.6919 with limits of agreement being -2.02 to 0.6936. Bland-Altman plot for cardiac output between cTD and NICaS™ showed a mean bias of -1.0386 ± 1.17 with limits of agreement being -3.34 to + 1.26. Percentage error for cardiac index and cardiac output were 64.78% and 64% respectively. Polar plot analysis showed an angular bias of 6.32° with radial limits of agreement being -8.114° to 20.75° for cardiac index and angular bias of 5.6682° with radial limits of agreement being -9.1422° to 20.4784° for cardiac output. Conclusion: NICaS™ demonstrated a good trending ability for both CI and CO. However, NICaS™ derived parameters are not interchangeable with the values derived from continuous thermodilution technique.
No takes yet. Share an insight, caveat, or question.
Singh et al. (2020) conducted an observational in Post-operative cardiac surgery (n=23). Non-Invasive Cardiac System (NICaS) vs. Continuous thermodilution technique (cTD) using pulmonary artery catheter was evaluated on Correlation of cardiac index between continuous thermodilution and NICaS (r = 0.765, 95% CI 0.70 to 0.82, p=<0.0001). Non-Invasive Cardiac System (NICaS) demonstrated a strong positive correlation with continuous thermodilution for measuring cardiac index (r = 0.765), but the parameters are not interchangeable due to wide limits of agreement and a 64% percentage error.