This is the authors' abstract. We don't add key points for this paper.
We describe the capability to measure high-resolution neutron powder diffraction data to a pressure of at least 8 kbar. We have used the HRPD instrument at the ISIS neutron source and a piston-cylinder design of pressure cell machined from a null-scattering titanium zirconium alloy. Data were collected under hydrostatic conditions from an elpasolite perovskite La2NiMnO6; by virtue of a thinner cell wall on the incident-beam side of the cell, it was possible to obtain data in the instrument's highest resolution back-scattering detector banks up to a maximum pressure of 8.5 kbar.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Bull et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: