This is the authors' abstract. We don't add key points for this paper.
The development of enzyme-free flexible glucose sensors is essential for advancing diabetes management, improving blood glucose control, and reducing the burden of monitoring. Herein, a simple two-step method was used to synthesize Cu-MoS2for glucose electrochemical sensing, which was then assembled onto flexible carbon cloth to form a flexible sensor (signed as Cu-MoS2). The Cu-MoS2exhibits a sensitivity of364~μA/(mM·cm2) over a wide linear range of1~μM–20 mM. It is reliable, reproducible, and electrochemically stable, with high specificity for glucose. After bending and stretching the Cu-MoS2, its sensing performance was re-evaluated. The results showed that Cu-MoS2retained excellent sensitivity, interference resistance, and stability, indicating its potential to meet the real-time blood glucose monitoring requirements of flexible sensors. Density generic function theory (DFT) calculations showed that the introduction of Cu into MoS2can enhance the adsorption and activation of glucose molecules, which is conducive to the oxidative dehydrogenation steps of –CHO and –OH in glucose, thereby improving the glucose sensing performance. The preparation method for the flexible sensor proposed in this work provides valuable technical guidance and shows promising potential for the development of flexible blood glucose sensors.