Answers are grounded in this paper and cite their sources.
Conventional approaches to CT dosimetry are inadequate because they fail to evaluate the risk to the patient. A risk related approach to CT dosimetry is developed which explicitly takes into account the non-uniform dose distribution in the body and the relative sensitivities of different organs and tissues. The principal radiological risks to patients undergoing CT examinations are the stochastic processes of carcinogenesis and genetic effects. Radiation risk estimates have been obtained for an EMI 5005 CT scanner by measuring the mean organ doses in a Rando phantom for CT studies on the head, chest, abdomen and pelvis. The application of these risk estimates to the population served by this CT scanner indicates that during the lifetime of the CT scanner, approximately 50,000 patients will undergo CT scanning and 60% will result in a positive diagnosis. The radiation detriment is estimated to be about 1 induced cancer and a negligible genetic effect. The radiation detriment is considerably smaller than the total detriment associated with contrast material used in 76% of the CT studies.
No takes yet. Share an insight, caveat, or question.
Huda et al. (1985) studied this question.