Key result
POC analysis of urine NT-proBNP proves feasible and stable for 96 hours.
Why the study?
While bedside serum NT-proBNP point of care testing is established, the analytical capabilities and room-temperature stability of NT-proBNP point of care testing in urine samples required investigation.
Is the cobas h232 point of care system feasible and accurate for determining NT-proBNP in urine samples?
Observational (n=104)
Is the cobas h232 point of care system feasible and accurate for determining NT-proBNP in urine samples?
Point-of-care testing for urine NT-proBNP is feasible and the biomarker is stable at room temperature for at least 96 hours, offering a potential non-invasive diagnostic tool once detection ranges are adjusted.
Answers are grounded in this paper and cite their sources.
May enable non-invasive urine NT-proBNP POC testing; leaves open clinical validation after detection-range adjustment.
INTRODUCTION: The diagnostic potential of urine NT-pro-BNP has been studied recently and shows great promise, especially for children. Bedside serum-NT-proBNP determination with point of care testing (POCT) is established in adults. This study investigated the analytical capabilities of NT-proBNP POCT measurements for urine-samples. Furthermore, the stability of this biomarker in urine samples at room-temperature was explored. METHODS: 104 urine samples were analysed, partly with double and triple determinations and compared to the established Elecsys system. To investigate the stability of NT-proBNP in urine, five samples were stored at room temperature and analysed 12-hourly over 96 h. RESULTS: of 0.84. The stability of urine NT-proBNP was shown with correlation coefficients between 0.97 and 1 (p < 0.001). CONCLUSION: Analysing NT-ProBNP in urine with cobas h232 POC system is feasible, but the detection range needs to be adjusted. As urine NT-proBNP proved stable for at least 96 h even postal shipping would be possible. Adjusting this system's measuring range could help introduce a new, non-invasive, pain-free parameter.
No takes yet. Share an insight, caveat, or question.
Müller et al. (2022) reported an observational. cobas h232 POC system vs. Elecsys system was evaluated on Stability of urine NT-proBNP (p=<0.001). Analyzing NT-ProBNP in urine with the cobas h232 POC system is feasible and stable for at least 96 hours (correlation coefficients between 0.97 and 1, p<0.001).