Key result
Protein A/G reliably detects IgG-mediated Atellica TnIH interference, outperforming highly variable PEG precipitation.
Why the study?
Macrotroponin may cause diagnostic confusion by elevating troponin results, leading to recommendations for clinical laboratories to investigate interference, but the applicability of detection methods for the Siemens Atellica troponin I assay needed evaluation.
Does Protein A/G IgG-depletion or PEG precipitation reliably detect macrotroponin interference with the Siemens Atellica high sensitivity troponin I assay?
Observational (n=85)
Does Protein A/G IgG-depletion or PEG precipitation reliably detect macrotroponin interference with the Siemens Atellica high sensitivity troponin I assay?
IgG depletion using Protein A/G can reliably establish IgG-mediated macrotroponin interference with the Atellica TnIH assay, whereas PEG precipitation results are highly variable and difficult to interpret.
Answers are grounded in this paper and cite their sources.
Urges caution interpreting PEG results for macrotroponin with Atellica TnIH; leaves open standardization of Protein A/G depletion.
OBJECTIVES: Macrotroponin refers to circulating immunoglobulin-bound cardiac troponin species that may elevate troponin results in patients with or without myocardial injury, causing diagnostic confusion. Clinical laboratories have been recommended to provide a service for troponin interference investigation. We evaluated the applicability of a Protein A/G IgG-depletion procedure as well as polyethylene glycol (PEG) precipitation for detecting macrotroponin interference with the Siemens Atellica troponin I (TnIH) assay. METHODS: Troponin I, IgG, and albumin (internal standard) were measured (Atellica) on the neat and treated plasma to calculate recovery. Reference samples with TnIH ranging from < 1x to > 1000x times the 99th percentile were selected to verify expected recovery. To minimize likelihood of macrotroponin in the reference group, samples with elevated results were only included if recent acute changes in TnIH was documented. 40 samples were used for the IgG-depletion method and 20 for PEG precipitation. 25 samples from patients with unexplained elevation in TnIH were assessed by IgG-depletion. RESULTS: 38 of 40 reference group recoveries exceeded 70 % (median 91 %, IQR 15 %, max 129 %) in the IgG-depletion group consistent with literature on other assays. Specimens from patients with incongruent clinical picture had IgG-depletion recovery median of 11 % (IQR 14 %, max 37 %). PEG-precipitation showed large variation (median 103 %, IQR 89 %, max 227 %). CONCLUSIONS: IgG depletion using Protein A/G can reliably establish IgG-mediated interference with Atellica TnIH. PEG precipitation results are difficult to interpret likely due to matrix effects, especially at values closer to the 99th percentile.
No takes yet. Share an insight, caveat, or question.
Karin et al. (2025) conducted an observational in Macrotroponin interference (n=85). Protein A/G IgG-depletion procedure and polyethylene glycol (PEG) precipitation was evaluated on Troponin I recovery after treatment. IgG depletion using Protein A/G reliably established IgG-mediated interference with Atellica TnIH, whereas PEG precipitation showed large variation.
Synapse has enriched 10 related papers. Consider them for comparative context: